[3.I] Specific safety or performance claims made for the device:<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 15 of 213<br />

introduction<br />

Manufacturer’s statement:<br />

For all risk applications or any use with patients, where falling would lead to very dangerous situations (for example after<br />

hip replacement operations, patients with neurologic problems, patients connected to invasive probes, maximum load<br />

tests, etc.) a fall prevention system like a safety arch with chest harness <strong>and</strong> stop-function or an unweighting system (for<br />

example the h/p/cosmos airwalk) is obligatory for such applications. Automatic stop of the <strong>running</strong> belt in the moment a<br />

fall occurs. The chest belt secures the subject <strong>and</strong> prevents falling over to the front. The chest belt straps do not pinch or<br />

restrict the athlete. The chest belt is connected via harness with the emergency stop at the crossbar of the safety arch. If<br />

there is more than approx. 8 kg / 17.6 lbs traction pulling on the harness, the switch will stop the <strong>running</strong> belt<br />

immediately. The length of the harness is individually adjustable.<br />

Unweighting systems as an accessory (for example the h/p/cosmos airwalk) can be used for body weight support of the<br />

patient, in case the patient should not bear his full body weight on his legs, for example after hip operation.<br />

A safety stop cord type (e.g. safety lanyard with magnetic switch) can be attached to the subject. The safety stop cord<br />

type shall limit the range of motion of the subject to the 70% of the front part of the <strong>running</strong> deck length.<br />

The actuator cord shall have a suitable length so that the actuator releases/triggers at a position not more than 70 % of<br />

the length of the <strong>running</strong> surface.<br />

In case the subject cannot catch up with the speed <strong>and</strong> is falling behind to the rear 30% of the <strong>running</strong> deck length, or in<br />

case the subject is falling, the safety stop cord type will trigger a stop function of the treadmill <strong>and</strong> will stop the <strong>running</strong><br />

belt.<br />

Caution! The safety stop cord type can not prevent from falling <strong>and</strong> is therefore not as safe as the safety arch with chest<br />

belt <strong>and</strong> fall prevention or the unweighting system!<br />

[3.J] Emergency buttons:<br />

The big <strong>and</strong> well visible emergency-stop-button is st<strong>and</strong>ard at all models, placed at the h<strong>and</strong>rail or UserTerminal for<br />

easy access. An additional emergency-stop-button can be retro-fitted at all h/p/cosmos <strong>running</strong> <strong>machine</strong>s, for example<br />

to be duplicated on the opposite site h<strong>and</strong>rail.<br />

Extract from the norm EN 957-6:<br />

The push-button type or pull cord safety stop actuator shall be in a reachable position in front of the user <strong>and</strong> in an area<br />

of ± 180 mm parallel to the centreline of the <strong>running</strong> surface.<br />

If the push-button type safety stop actuator is outside this area it shall be duplicated on the opposite side except when a<br />

pull-cord actuator is provided.<br />

h/p/cosmos can supply a second <strong>and</strong> additional push button emergency stop button <strong>and</strong>/or a pull-cord actuator (safety<br />

lanyard with magnetic stop) <strong>and</strong>/or a safety arch with chest belt, harness <strong>and</strong> fall stop function.<br />

At all models of the category medical the main push button emergency-stop is simultaneously an emergency-off, that<br />

means by pushing this button the power supply of the device gets cut off.