[10.I] Error messages / Error codes<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 147 of 213<br />

trouble shooting<br />

Interferences <strong>and</strong> error codes might be caused by problems with the power supply or because of not enough<br />

maintenance (lubrication of the <strong>running</strong> surface, etc.).<br />

N Check voltage supply. Do not use extension cords or multi-way connectors. Connect the <strong>machine</strong> direct to the<br />

socket in the wall. Each <strong>machine</strong> should have an individual circuit.<br />

N Check mechanical parts for any malfunction. Check also for towels or other things, which might interfere the drive<br />

system.<br />

N Check if any light barrier are dusty or not correctly adjusted.<br />

N Check lubrication of the <strong>running</strong> belt <strong>and</strong> lubricate if required.<br />

N Also contact problems at connectors (loose connections) caused by vibrations can result in malfunctions. So<br />

please check cables <strong>and</strong> connectors for loose connections.<br />

The device has a self-diagnosis which recognizes some errors <strong>and</strong> shows their error messages on the display of the<br />

MCU UserTerminal (operating terminal) or the frequency changer / inverter (inside the device).<br />

[10.I1] Devices with UserTerminal<br />

Start: Running belt is not moving. One of the mode LED � of operation is flashing: (manual, profile, cardio, test)<br />

[10.I1] Message Description Explanation / Display / Activity<br />

E01 Oil interval due = „E 01“ ! „OIL“ flashing<br />

§ $ % „HELP“ flashing<br />

Running <strong>machine</strong> is still functioning<br />

N Every 1000 km (interval is adjustable) lubrication service is<br />

due. See chapter "lubrication of the <strong>running</strong> belt". Lubricate the<br />

belt.<br />

N Attention: Running <strong>machine</strong> has no sensor for oil. Lubrication<br />

only according to operation manual. Error message does not<br />

disappear automatically.<br />

N Set back error message (option 01)<br />

E02 Service interval due<br />

= „E 02“ ! „Service“ flashing<br />

or<br />

§ $ % „HELP“ flashing<br />

Safety inspection due<br />

Running <strong>machine</strong> is still functioning<br />

N Every 5000 km (interval is adjustable) a general service<br />

(interior cleansing, checking the driving belt, <strong>running</strong> belt, etc.)<br />

is due.<br />

N Every 12 months a electric saftey inspection is due. See<br />

chapter maintenance in this manual.<br />

N Contact authorized service engineer<br />

N Set back error message (option 01)