optional settings<br />
[8.B1] Continuation… N Running <strong>machine</strong> series h/p/cosmos quasar with<br />
special speed 45 km/h:<br />
Speed Calibration Value: 2.874<br />
Inverter Frequency: 125 Hz<br />
N Running <strong>machine</strong> series h/p/cosmos pulsar 3p with 40 km/h:<br />
Speed Calibration Value: 2.874<br />
Inverter Frequency: 110 Hz<br />
N Running <strong>machine</strong> series h/p/cosmos pulsar 3p with<br />
special speed 45 km/h:<br />
Speed Calibration Value: 2.874<br />
Inverter Frequency: 125 Hz<br />
OP35 Oil Interval<br />
The distance interval for indication of "OIL" service required can be<br />
adjusted by means of option 35.<br />
St<strong>and</strong>ard = 1000 km for treadmills, 100 km for ladder-ergometers. OIL<br />
intervals are indicated with error code "E01" "OIL" every 1000 km.<br />
§ reports: oil interval in km (St<strong>and</strong>ard = 1.000)<br />
� km is flashing<br />
Adjustment range: OFF, 100 ... 5.000 km<br />
Do not change the interval without prior communication with<br />
h/p/cosmos. For the models h/p/cosmos venus, saturn und orbiter,<br />
JAEGER LE 580 <strong>and</strong> bigger as well as COSMED T200 <strong>and</strong> bigger<br />
option number 35 is not adjustable, as the <strong>machine</strong>s are equipped with<br />
an automatic oil pump <strong>and</strong> a tank.