maintenance & safety inspections<br />

N Clean device <strong>and</strong> engine compartment/internal space. Remove dust <strong>and</strong> dirt from the cooler openings of the<br />

fan housing <strong>and</strong> the cooling air ducts of the drive motor, as well as from the ventilation slots <strong>and</strong> the cover of<br />

perforated plate of the frequency inverter.<br />

N Attach hazard notes completely <strong>and</strong> well legible according to the instructions. Check the warning signs,<br />

protective covers, step platform, engine hood <strong>and</strong> plastic-cover installation channel in the engine compartment<br />

for existence <strong>and</strong> for damage <strong>and</strong> replace as necessary.<br />

N All protective-resistance connectors, which can be reached from the outside, must examined for their correct<br />

value <strong>and</strong> correct labelling.<br />

N Check capture slot at the rear end <strong>and</strong> at reverse belt rotation in the front as well <strong>and</strong> adjust as necessary.<br />

Capture slot < 8 mm. According to regulation draft 60601-2-xx © IEC:200X 62D/479/NWIP 2003-05-18 <strong>and</strong><br />

according to EN 957-1. See chapter “test-finger“, which has a diameter of 9.5mm (+0.1) at the front.<br />

[9.C2] Protective Earth Resistance RPE measurement<br />

[9.C3] Isolation Resistance RISO measurement<br />

N Resistance between housing <strong>and</strong> ProtectiveEarth-<br />

Connection.<br />

N The low resistance pass is to be controlled according to<br />

VDE 0701/0702 (sports <strong>and</strong> fitness <strong>machine</strong>s) or VDE<br />

0751/IEC 60601-1 (medical devices) by the protectiveresistance<br />

measurement with a measuring device for the<br />

protective-earth-resistance measurement.<br />

N The connecting lead is to be moved while measuring for<br />

at least 5 sec. If the resistance changes hereby it is highly<br />

probable that the cable or the connectors have been<br />

damaged. In this case the cable has to be replaced <strong>and</strong><br />

the device has to be repaired immediately.<br />

N Resitance between „briged“ L+N <strong>and</strong> ProtectiveEarth-<br />

Connection.<br />

N Make sure that all isolations, that are under stress of the<br />

mains voltage, are being seized. All switches <strong>and</strong><br />

contactors should be connected.<br />

N The measuring is to be performed with measuring<br />

devices for the isolation-resistance measurement<br />

according to VDE 0701/0702 (sports <strong>and</strong> fitness<br />

<strong>machine</strong>s) or VDE 0751/IEC 60601-1 (medical devices).<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 124 of 213