© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 149 of 213<br />

trouble shooting<br />

[10.I1] N Clean the speed sensor (light barrier <strong>and</strong> disk, where<br />

applicable) <strong>and</strong> check the adjustment of the sensor according<br />

to the instruction.<br />

N Check if the safety-, acceleration- <strong>and</strong> deceleration time (of the<br />

frequency inverter <strong>and</strong>/or MCU) is higher than the highest<br />

acceleration level (e.g. sprints with very high acceleration<br />

levels, respectively at using the SPRINT-function if available).<br />

N Erase the message with "OPTION 01". In case the message<br />

appears again, contact authorized service engineer.<br />

N Check also Administrator OPTION 48<br />

N Check plug no. S6 on MCU <strong>and</strong> analogue voltage 0 ... 10 volts<br />

(for 0 ... max. speed) from MCU board to inverter drive.<br />

N Check plug no. S6 on MCU <strong>and</strong> “inverter on – signal / RFR”<br />

from MCU board to inverter drive<br />

E31 + tolerance aberration speed sensor<br />

The measured speed-value<br />

(instantaneous value) is at least the in<br />

Check mains power supply to frequency inverter<br />

= „E 31“ ! „INCR“ flashing<br />

§ $ % „HELP“ flashing<br />

administrator-option 12 set percentage<br />

BIGGER than the calculated value<br />

(debit-value)<br />

Error search see E30.<br />

E32 - tolerance aberration speed sensor<br />

The measured speed-value<br />

(instantaneous value) is at least the in<br />

= „E 32“ ! „INCR“ flashing<br />

§ $ % „HELP“ flashing<br />

administrator-option 12 set percentage<br />

LOWER than the calculated value (debitvalue)<br />

Error search see E30.