[7.B13] 2 Graded test Continuation ...<br />
3 Conconi test<br />
e.g. for determination of anaerobic threshold via<br />
heart rate curve of the subject<br />
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />
© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 78 of 213<br />
operation<br />
In an already pre-defined step-test it is possible, to set the preprogrammed<br />
pause-times individually during the test, i.e. the<br />
pause can be prolonged or ended immediately.<br />
End pause manually:<br />
Once the test profile has reached the pause mode <strong>and</strong> the<br />
<strong>running</strong> belt has come to halt, you can select the next program<br />
step immediately with START – either with warning sound<br />
signals at the beginning by pressing START once or without by<br />
pressing START twice. This is done independently of the predefined<br />
pause-times <strong>and</strong> has no influence on the following run<br />
of the program, i.e. the following program steps remain<br />
unaffected. By pressing the START button the next speed<br />
step scheduled is started. The pause can also be ended with<br />
the „+“ button. In this case the speed set depends on the point<br />
of release of the „+“ button.<br />
Prolong pause:<br />
Once the test profile has reached the pause mode <strong>and</strong> the<br />
<strong>running</strong> belt has come to halt, it is possible to freeze pause<br />
with the „-“ button.<br />
= indicates: „PAUS“<br />
To end pause you can either press START or „+“ - see end<br />
pause manually.<br />
Endurance test (max. heart rate test)<br />
St<strong>and</strong>ard load profile:<br />
N Starting speed: 8.0 km/h, must be changed according the<br />
conditions of the subject<br />
N Circuit (lap length): 200m (can be changed)<br />
N Increment: 0.5 km/h (can be changed)<br />
STOP must be activated manually by the medical doctor when<br />
the subject is fully exhausted.