running machine - H-P-COSMOS Sports and Medical
running machine - H-P-COSMOS Sports and Medical
running machine - H-P-COSMOS Sports and Medical
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
installation & commissioning<br />
Devices for <strong>Sports</strong> <strong>and</strong> Fitness manufacturing before manufacturing date 12/2007<br />
The <strong>running</strong> <strong>machine</strong> can be switched on directly at the expulsion fuse at the FrontTerminal (frontal section below<br />
the hood) to the condition "st<strong>and</strong>-by mode" (if the emergency-stop has been activated a flashing notice „PULL<br />
STOP“ appears on the display.).<br />
Expulsion fuse <strong>and</strong><br />
main switch 2- or 3-pole<br />
[6.D3] Switching the device off<br />
ON / OFF button<br />
with control lamp<br />
ON / OFF button<br />
until 2003<br />
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />
© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 57 of 213<br />
Main switch from construction<br />
size <strong>running</strong> surface<br />
200...300/75...125cm onwards<br />
N The ON-/OFF-intervals must not be shorter than 1-2 minutes (at models with 3-phase<br />
connection: 2-3 minutes). Otherwise it could lead to interference in the adjustment of<br />
the motor or to the failure of the backup.<br />
N The models for medical application with the isolation transformer have an inrush-<br />
current limiter (surge guard). Too short switch on / switch off intervals will lead to a<br />
deactivation of the inrush current limiter <strong>and</strong> results in an overload of the circuit fuse.<br />
N For the professional use, where the devices are often being used daily, we recommend<br />
to switch on the devices in the morning, <strong>and</strong> leave it in the st<strong>and</strong>-by mode during the<br />
day.<br />
Switch off the <strong>running</strong> <strong>machine</strong> by pressing the black key (until 2003 red “O” key) at the UserTerminal (some<br />
models at the front of the device). The indicator within the key is extinguished. At some models built before<br />
12/2007 the location of the ON-/OFF-key could also be the UserTerminal.<br />
Devices for <strong>Sports</strong> <strong>and</strong> Fitness manufacturing before 12/2007<br />
The <strong>machine</strong> can be switched off directly with the explusion fuse at the FrontTerminal (frontal section below the<br />
hood).