LCD Display: ALERT NEW DOSE IS LESS THAN CURRENT DOSEAlert: Unable to program a lower number of doses than number of doses alreadyinfusedCause: While the Variable or Intermittent therapy is in progress, a lower numberof doses to be delivered is entered than the number already infused.Resolution: Wait until current therapy in progress is complete, or “PAUSE” pumpand, using the REPEAT Rx function, modify the number of doses.Note: Selecting to repeat the therapy in progress will terminate the currenttherapy.TroubleshootingLCD Display:Alert:Cause:Resolution:LCD Display:Alert:Cause:Resolution:LCD Display:Alert:Cause:Resolution:LCD Display:Alert:Cause:Resolution:ALERT NO BOLUS DOSE DEFINEDUnable to enter bolus interval or # of bolus/hr because no bolus value hasbeen programmedEntry of a value other than zero for the Bolus Interval or #Bolus per hourfield when no bolus value has been programmed.Enter bolus value before entering bolus interval or # of boluses/hr.ALERT NO KEYPAD ACTIVITYNo Keypad ActivityThe pump has been displaying a non-running screen for two minuteswithout any keys being pressed. The alert will recur every 2 minutes aslong as there are no key presses.Complete the activity for the displayed menu. The pump may also beturned off, however, all programming that was not complete will bediscarded.If the pump was running and the Options menu displayed when the alertoccurred, the run screen for the therapy will be displayed when the alert iscleared. Any changes to the Options menu are discarded and will have tobe re-entered.ALERT NO THERAPY ENABLEDNo Therapy EnabledAttempting to accept the Enable Therapy menu with all therapies set to“OFF”. At least one therapy must be enabled or “ON” for menu to beaccepted.Enable the therapy or therapies that are to be active for the pump.ALERT NOT ALLOWED, DOSE IN PROGRESSChange is not allowed while a dose is in progressModification to the current field is not allowed while the dose is inprogress.Wait until the dose in progress completes before modifying field.127

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!