Adapter eliminator and remove one C-Cell battery.) Observe that the AuxiliaryAlarm System is activated, with the Red LED and Beeper on constant (notintermittently beeping). With the battery still out of the pump, press the ON/OFFbutton to turn off the auxiliary alarm. Replace the battery in the batterycompartment and turn on the pump. Resume the therapy and confirm that theinfusion continues where it stopped due to Auxiliary Alarm occurrence.4 SET BIOMED CLINICIAN MENU ITEMS:a Turn the pump Off and back On. While pump is performing Self Tests confirm thatthe pump is set to the correct Date and Time. If date and time are not correct,program the correct date and time from the Biomed/Clinician menu.b Enter the Biomed/Clinician Menu using the Clinician Code.c Set the Lock Level to OFF in the CHANGE LOCK menu.d Clear the Pt DATA from the pump.e Observe the Pump Maintenance Date, top of the Biomed/Clinician Menu, to insureadequate time for the next use by patient.f Sign, date and return the Check List to your record file. Pumps that pass can beused for new patients. Pumps that fail are to be returned for service.150