In CMS-enabled pumps, the graphic bar will toggle with Line Pressure (“L Press”)which displays the current patient side (downstream) pressure in the administrationset in millimeter of mercury (mmHg) units.3 The third line displays the total volume infused and, if mg or mcg is the selected unit, this fieldwill toggle to display the total (mg or mcg) infused as well.4 The last line displays the remaining time to the completion of the infusion toggled with theremaining amount to be infused.5 In addition to the display screen giving information regarding the status of the infusion, thegreen LED light blinks whenever the pump is infusing.The information on the run screen will alternate between a four-line display and a large-fontdisplay of each line of information on the screen.Continuous Information ScreensInterrupting an InfusionThe infusion can be paused at any time by pressing the “PAUSE” key. The yellow “Standby”LED light blinks, and the following screen displays:Pause Menu ScreenContinuous Therapy51