IntroductionChapter1IntroductionThe
- Page 9 and 10:
• Retained memory features that a
- Page 11 and 12:
CAUTIONS• U.S. federal law restri
- Page 13 and 14:
• This equipment generates, uses,
- Page 15 and 16:
Serial Number LabelIntroductionBatt
- Page 17 and 18:
Preparing Medication for Infusion/C
- Page 19 and 20:
Opening and Closing the Pump DoorTh
- Page 21 and 22:
Air DetectorNote: Open theslide cla
- Page 23 and 24:
Installing the Batteries and Using
- Page 25 and 26:
IMPORTANTINFORMATIONWhenever an ext
- Page 27 and 28: IntroductionLocking SleeveTwo-Pin A
- Page 29 and 30: Chapter2Learning Pump FeaturesLearn
- Page 31 and 32: LED Light IndicatorsThis “three-l
- Page 33 and 34: Chapter3Basics of ProgrammingThe Cu
- Page 35 and 36: whenever an alarm or error code occ
- Page 37 and 38: Access CodeClinician Access Code Sc
- Page 39 and 40: to all pump functions. Each level h
- Page 41 and 42: 8 PRINT PATIENT HISTORY: To print t
- Page 43 and 44: a. Time Format: You have the option
- Page 45 and 46: If only one therapy is programmed a
- Page 47 and 48: Chapter4Continuous TherapyContinuou
- Page 49 and 50: If the KVO rate is set to 0, the pu
- Page 51 and 52: 6 DONE?: If any changes are to be m
- Page 53 and 54: The following screen displays the C
- Page 55 and 56: detects in-line pressures at 400 mm
- Page 57 and 58: In CMS-enabled pumps, the graphic b
- Page 59 and 60: . If the incomplete therapy is to b
- Page 61 and 62: PCA TherapyChapter5PCA TherapyPCA o
- Page 63 and 64: a. MxRATE (Maximum Rate) limits: Us
- Page 65 and 66: using the numeric keys and, when co
- Page 67 and 68: Note. After the infusion has starte
- Page 69 and 70: d. To exit the shift total feature
- Page 71 and 72: pressures at 400 mmHg ±150 (8 ±3
- Page 73 and 74: This screen will continue to displa
- Page 75 and 76: When the program is accepted, the d
- Page 77: which therapy to repeat by moving t
- Page 81 and 82: At the following screen, the prescr
- Page 83 and 84: administration set was first connec
- Page 85 and 86: pump has been set into a locked con
- Page 87 and 88: Also, when the Air-In-Line sensor i
- Page 89 and 90: The screen offers the user three op
- Page 91: This screen displays:1 “INFUSION
- Page 96 and 97: Note: You may elect to skip the dos
- Page 98 and 99: 4 Release any clamps on the adminis
- Page 100 and 101: 3 CLR HRLY (Clear Hourly Totals): T
- Page 102 and 103: Reviewing or Changing the ProgramTo
- Page 104 and 105: Run Options Screena. Press the “R
- Page 106 and 107: Timing Rules when Interrupting Inte
- Page 108 and 109: 102
- Page 110 and 111: “ON.” Now, when you press the
- Page 112 and 113: “YES/ENTER’ to accept the new d
- Page 114 and 115: PrimeThe IV bag and administration
- Page 116 and 117: 1 LOCK: When the lock is set to “
- Page 118 and 119: Also, when the Air-In-Line sensor i
- Page 120 and 121: Interrupting an InfusionThe infusio
- Page 122 and 123: Note: Selecting “NEW PROGRAM” w
- Page 124 and 125: 118
- Page 126 and 127: If no air-eliminating filter is bei
- Page 128 and 129:
LCD Display:Alert:Cause:Resolution:
- Page 130 and 131:
Alert:Cause:Resolution:LCD Display:
- Page 132 and 133:
LCD Display:Alert:Cause:Resolution:
- Page 134 and 135:
LCD Display:Alert:Cause:Resolution:
- Page 136 and 137:
LCD Display:Alert:Cause:Resolution:
- Page 138 and 139:
LCD Display: ALARM EMPTY BATTERYAla
- Page 140 and 141:
LCD Display:Alarm:Cause:Resolution:
- Page 142 and 143:
ErrorCodeLCD DisplayedMessage34 Tur
- Page 144 and 145:
ErrorCodeLCD DisplayedMessage204 Tu
- Page 146 and 147:
Remote Bolus Cord (340-2005)This fi
- Page 148 and 149:
Pump Holster (340-0111)This accesso
- Page 150 and 151:
144
- Page 152 and 153:
Transport and StorageDuring transpo
- Page 154 and 155:
Procedure for Checking Pumps Betwee
- Page 156 and 157:
Adapter eliminator and remove one C
- Page 158 and 159:
ITEMOperating Time with Full Charge
- Page 160 and 161:
Percent Error, Curlin Pump infusing
- Page 162 and 163:
Percentage Error, Curlin Pump deliv
- Page 164 and 165:
Flow Rate (1 hourcalculation window
- Page 166:
Lock Setting TableThe following tab