ITEMOperating Time with Full ChargeBatteries:Internal Memory BatteryDisplayFlow RatesKVO RatesVolume AccuracyAdministration SetsFill Volume for Administration SetsUp Occlusion Detection TimeMicrobore tubing with occlusionbelow spikeSPECIFICATIONNominal 10 hours at a rate of 400 ml/hrNominal 30 hours at a rate of 125 ml/hrNominal 85 hours at a rate of 2 ml/hr5 years lifeGraphic LCD, 4 text fields, 16 characters perfield0.1 to 400 ml/hr0 to 10 ml/hr. Can be set at “0” in Continuous,TPN, Variable and Intermittent Therapies± 5% for volume delivery >1 ml under thefollowing conditions:• Temperature 55° to 90° F• Back pressure to 300 mmHg (5.8 PSI)±10% for volume delivery >1 ml under thefollowing conditions:• Temperature 45° to 55°F• Temperature 90° to 104°F• Back pressure to 750 mmHg (14.5 PSI)The maximum volume that may be infused underSingle Fault Conditions is 0.8 ml.Recommended administration set usage not toexceed 72* hours.*follow standards of practice for set changesProprietary <strong>Curlin</strong> <strong>Med</strong>ical non-DEHP, PVCadministration sets available in differentconfigurations.Nonfiltered microbore sets ± 4 ml.See individual administration set label for specificfill volumeRate0.1 ml/hr1 ml/hr10 ml/hr50 ml/hrTime to Alarm