Flow-StophandleAir DetectorSlot1 Insert blue “Tubing Guide” on rightside of pump with tubing in front ofthe door. The Blue arrow on thebackground surface of the pump headindicates direction and location for theblue Tubing Guide.2 Important: Center the tubinghorizontally over the pumping fingerscovering the yellow dot.3 Insert yellow “Integral Flow-Stop”on the left side of the pump. Theyellow arrow on the backgroundsurface of the pump head indicates thedirection and location for the “IntegralFlow-Stop.”CAUTION: Hold the “Flow-Stopby the small handle. DO NOTpress down on the top of the“Integral Flow-Stop” wheninserting it into the receptacle.This action could inadvertentlyopen the “Integral Flow-Stop” andallow fluid to flow through thetubing.4 Press the tubing into the air detectorslot gently with index finger.Figure 1.10 Top cutaway view of the <strong>Curlin</strong> <strong>Med</strong>ical pump with <strong>Curlin</strong> <strong>Med</strong>icalAdministration set in proper installation positionUp Stream Occlusion Alarm FeatureThe <strong>Curlin</strong> <strong>Med</strong>ical pump has an Up Stream Occlusion alarm feature, e.g., it detects low upstreampressure. This feature notifies you if an occlusion is present between the IV container andthe pump. If this alarm occurs, the display will read, “ALARM –UP OCCLUSION,” the alarmwill sound, the red LED light will blink, and the infusion will stop. This situation requires that youpress the “Pause” key, examine the administration set from the IV container to the pump forpotential causes of the occlusion. If a clamp is in place, open, or remove it; if the administrationset is kinked, straighten it. If indicated, close the slide clamp on the administration set and openthe door of the pump to verify that the set is correctly placed. Close the pump’s door and reopenthe slide clamp. When ready to begin infusing, select “RESUME” and then press the “RUN”key.The pump also detects excessive up stream pressure (pressure on the IV bag). The display willread “ALARM – HIGH UP PRESSURE,” the alarm will sound, the red LED light will blink, andthe infusion will stop. If this situation occurs, relieve the source of excessive pressure on the IVbag and when ready to begin infusing, select “RESUME” and then press the “RUN” key.Note: The up stream occlusion detection is disabled during priming.16
Installing the Batteries and Using External PowerThe <strong>Curlin</strong> <strong>Med</strong>ical pump can be powered solely with the use of two “C” size alkaline batteriesinstalled into the pump as shown in Figure 1.11. “C” size batteries are readily available at mostretail facilities that carry alkaline batteries. A second alternative for power is the portable,rechargeable Battery Pack, shown in Figure 1.12. The third alternative for power is to use the ACAdapter shown in Figure 1.13, allowing the pump to be connected to any approved, three-prongelectrical wall outlet.IntroductionThe “C” size batteries will last approximately 30 hours at a rate of 125 ml/hour and over 85 hoursat rates of 2 ml/hr or less. Factors that influence battery life are the number of times the screen isilluminated with a key press, the viscosity of the fluid being delivered, the age and temperature atwhich the batteries have been stored. To extend the life of the batteries, patients can use the ACadapter whenever they are stationary, e.g., at night.The pump has a “low battery” alert that beeps periodically with a message appearing to tell youwhen the power is low in the “C” size batteries and/or the battery pack. This alert gives you awindow of time to be sure you have back-up power (e.g. extra batteries, AC adapter, or arecharged battery pack.) At a continuous infusion rate of 25 ml per hour, you can expect at least a30 minute warning before your pump can no longer operate due to empty batteries. At higherinfusion rates your advance warning will be less.The user should be taught that whenever the “low battery” alert appears, to pause the therapy,turn the pump off, and to replace both “C” size batteries. Then, turn the pump back on, selectresume to continue the therapy.When the “empty battery” alarm occurs the pump will go into an alarm state and the infusion willstop. If this occurs, turn the pump off, replace both of the “C” size batteries, turn the pump on,and go to the “Resume” function to continue the therapy. All programming information isretained. If the “C” size batteries are empty and you want to connect the AC adapter to powerthe pump, turn the pump off first, plug in the AC adapter and then turn the pump back on.Note: If you are operating the pump solely on the battery pack or the AC Adapter (no “C” sizebatteries installed) and the power source is disconnected from the pump, the alarm willsound, the infusion will stop, and the display screen will be blank. Turn the pump off,reconnect the power source, turn the pump on, and resume the therapy. All programmedinformation is retained.When replacing batteries, pause any infusion in progress by pressing the “PAUSE” key,and then turn the pump off. Replace the batteries, and, when the pump is turned back on,select “RESUME” to continue the therapy from the point at which it was interrupted.17
- Page 1 and 2: CURLIN MEDICALUser’s Manualfor us
- Page 3 and 4: TABLE OF CONTENTSEntering or Changi
- Page 5 and 6: TABLE OF CONTENTSPrime.............
- Page 7 and 8: IntroductionChapter1IntroductionThe
- Page 9 and 10: • Retained memory features that a
- Page 11 and 12: CAUTIONS• U.S. federal law restri
- Page 13 and 14: • This equipment generates, uses,
- Page 15 and 16: Serial Number LabelIntroductionBatt
- Page 17 and 18: Preparing Medication for Infusion/C
- Page 19 and 20: Opening and Closing the Pump DoorTh
- Page 21: Air DetectorNote: Open theslide cla
- Page 25 and 26: IMPORTANTINFORMATIONWhenever an ext
- Page 27 and 28: IntroductionLocking SleeveTwo-Pin A
- Page 29 and 30: Chapter2Learning Pump FeaturesLearn
- Page 31 and 32: LED Light IndicatorsThis “three-l
- Page 33 and 34: Chapter3Basics of ProgrammingThe Cu
- Page 35 and 36: whenever an alarm or error code occ
- Page 37 and 38: Access CodeClinician Access Code Sc
- Page 39 and 40: to all pump functions. Each level h
- Page 41 and 42: 8 PRINT PATIENT HISTORY: To print t
- Page 43 and 44: a. Time Format: You have the option
- Page 45 and 46: If only one therapy is programmed a
- Page 47 and 48: Chapter4Continuous TherapyContinuou
- Page 49 and 50: If the KVO rate is set to 0, the pu
- Page 51 and 52: 6 DONE?: If any changes are to be m
- Page 53 and 54: The following screen displays the C
- Page 55 and 56: detects in-line pressures at 400 mm
- Page 57 and 58: In CMS-enabled pumps, the graphic b
- Page 59 and 60: . If the incomplete therapy is to b
- Page 61 and 62: PCA TherapyChapter5PCA TherapyPCA o
- Page 63 and 64: a. MxRATE (Maximum Rate) limits: Us
- Page 65 and 66: using the numeric keys and, when co
- Page 67 and 68: Note. After the infusion has starte
- Page 69 and 70: d. To exit the shift total feature
- Page 71 and 72: pressures at 400 mmHg ±150 (8 ±3
- Page 73 and 74:
This screen will continue to displa
- Page 75 and 76:
When the program is accepted, the d
- Page 77:
which therapy to repeat by moving t
- Page 80 and 81:
The time format (12 hr AM/PM or 24
- Page 82 and 83:
time period using the numeric keys,
- Page 84 and 85:
7 To continue priming, press the PR
- Page 86 and 87:
Note: Whatever Down Ramp time was a
- Page 88 and 89:
1 The first field displays the “R
- Page 90 and 91:
. If the incomplete therapy is to b
- Page 95 and 96:
IMPORTANTINFORMATIONDuring the dela
- Page 97 and 98:
10 DONE?: If there are any changes
- Page 99 and 100:
The following screen displays the I
- Page 101 and 102:
This message alerts you to the Air-
- Page 103 and 104:
Intermittent Information ScreensInt
- Page 105 and 106:
Intermittent Dosing ScheduleThe dos
- Page 107 and 108:
This screen displays:1 “INFUSION
- Page 109 and 110:
CHAPTER8Variable TherapyThis therap
- Page 111 and 112:
During the delay start period, if t
- Page 113 and 114:
5 NEXT?: If there are any changes t
- Page 115 and 116:
7 To continue priming, press the PR
- Page 117 and 118:
4 DN Occlu (Down Occlusion): This f
- Page 119 and 120:
Variable Run Screen 2This screen di
- Page 121 and 122:
stop the auto-advance function by p
- Page 123 and 124:
The pump will then display the prog
- Page 125 and 126:
TroubleshootingChapter9Troubleshoot
- Page 127 and 128:
LCD Display:Alert:Cause:Resolution:
- Page 129 and 130:
If it is not necessary to preserve
- Page 131 and 132:
Resolution:LCD Display:Alert:Cause:
- Page 133 and 134:
LCD Display: ALERT NEW DOSE IS LESS
- Page 135 and 136:
LCD Display:Alert:Cause:Resolution:
- Page 137 and 138:
ALARMSAlarms are conditions that re
- Page 139 and 140:
LCD Display:Alarm:Cause:Resolution:
- Page 141 and 142:
ERROR CODESIf a pump goes into an e
- Page 143 and 144:
ErrorCodeLCD DisplayedMessage162 St
- Page 145 and 146:
ChapterAccessories10There are a num
- Page 147 and 148:
AccessoriesFigure 10.3 Safety Shell
- Page 149 and 150:
Keypad CoversThese accessories are
- Page 151 and 152:
Chapter11Cleaning, EnvironmentalCon
- Page 153 and 154:
1 PARTIES COVERED BY THIS WARRANTY:
- Page 155 and 156:
d Program the following CONTINUOUS
- Page 157 and 158:
Chapter12TechnicalSpecificationsThe
- Page 159 and 160:
ITEMDown Occlusion Detection TimeMi
- Page 161 and 162:
Curlin pump inusing at .1 ml/hr w i
- Page 163 and 164:
Percent Error, Curlin Pump infusing
- Page 165 and 166: