R&S ® <strong>WinIQSIM2</strong><br />

Graphical User Interface<br />

Baseband Block<br />

Note: This function is not affected by the "Preset" function.<br />

Remote command:<br />

​[:​SOURce]:​BB:​DM:​SETTing:​STORe:​FAST​ on page 261<br />

​[:​SOURce]:​BB:​ARBitrary:​MCARrier:​SETTing:​STORe:​FAST​ on page 223<br />

refer to the description of frimware options for the syntax of the corresponding SCPI<br />

commands<br />

Data Source<br />

The data source is selected and set in the "Data Source" section. The parameters offered<br />

depend on the data source selected. The More button opens a submenu for calling the<br />

data editor and the file manager.<br />

Data Source - Digital Modulation<br />

Selects the data source (see ​chapter 5.10.1, "Data and Signal Sources in Baseband",<br />

on page 48).<br />

You may choose from the following data sources:<br />

"All 0, All 1"<br />

"PRBS, PRBS<br />

Type"<br />

0 data or 1 data is internally generated.<br />

PRBS data in accordance with the IUT-T with period lengths between<br />

29-1 and 223-1 are internally generated.<br />

The length is selected in the "PRBS Type" input box.<br />

Remote command:<br />

​[:​SOURce]:​BB:​DM:​PRBS[:​LENGth]​ on page 234<br />

"Pattern"<br />

A user-definable bit pattern with a maximum length of 64 bits is internally<br />

generated.<br />

The bit pattern is defined in the "Pattern" input box.<br />

"Data List ...,<br />

Select Data"<br />

Data lists will be used.<br />

Data lists can be generated internally in the data editor or externally.<br />

Data lists are selected in the "File Select" window, which is called by<br />

means of the "Select Data" button.<br />

The "File" "Manager" is used to transmit external data lists to the R&S<br />

Signal Generator, and can be called within every "File" Select window<br />

by means of the "File" "Manager" button.<br />

Remote command:<br />

​[:​SOURce]:​BB:​DM:​DLISt:​SELect​ on page 257<br />

Remote command:<br />

​[:​SOURce]:​BB:​DM:​SOURce​ on page 235<br />