R&S ® <strong>WinIQSIM2</strong><br />

Remote Control Commands<br />

SOURce:BB:DM Subsystem<br />

Parameters:<br />

<br />

Example:<br />

float<br />

The range of values depends on the symbol rate that is set (see<br />

data sheet) and the maximum deviation is 10 MHz.<br />

Range: -40E6 to 40E6<br />

Increment: 0.5<br />

Default unit: Hz<br />

BB:DM:FORM FSKV<br />

selects Variable FSK modulation.<br />

BB:DM:FSK:VAR:TYPE FSK16<br />

selects 16FSK modulation.<br />

BB:DM:FSK:VAR:SYMB0:DEV 135000<br />

sets the frequency deviation of the least significant symbol to 135<br />

kHz.<br />

<strong>Manual</strong> operation: See "Deviation xxxx - Digital Modulation" on page 71<br />

[:SOURce]:BB:DM:FSK:VARiable:TYPE <br />

The command selects the modulation type for Variable FSK.<br />

Parameters:<br />

<br />

Example:<br />

FSK4 | FSK8 | FSK16<br />

*RST:<br />

FSK4<br />

BB:DM:FORM FSKV<br />

selects Variable FSK modulation.<br />

BB:DM:FSK:VAR:TYPE FSK16<br />

selects 16FSK modulation.<br />

<strong>Manual</strong> operation: See "FSK Type - Digital Modulation" on page 71<br />

7.8.1.5 Power Ramp<br />

[:​SOURce]:​BB:​DM:​PRAMp:​ATTenuation​..................................................................247<br />

[:​SOURce]:​BB:​DM:​PRAMp:​FDELay​.........................................................................248<br />

[:​SOURce]:​BB:​DM:​PRAMp:​RDELay​.........................................................................248<br />

[:​SOURce]:​BB:​DM:​PRAMp:​SHAPe​..........................................................................248<br />

[:​SOURce]:​BB:​DM:​PRAMp:​TIME​.............................................................................249<br />

[:​SOURce]:​BB:​DM:​PRAMp[:​STATe]​.........................................................................249<br />

[:SOURce]:BB:DM:PRAMp:ATTenuation <br />

The command sets the level attenuation for signal ranges that are flagged with level<br />

attribute attenuated by the LEV_ATT control signal.<br />

<strong>Software</strong> <strong>Manual</strong> 1008.0858.42 ─ 10<br />

247