MI 611-185 – June 1996<br />

Appendix A: Text Data Formats<br />

CAL AUTO ACCEPT=OFF<br />

CAL AUTO SAVE=OFF<br />

CAL UNIT LOCK=OFF<br />

BACKGROUND CORRECTION (PID)=ON<br />

BACKGROUND CORRECTION (FID)=ON<br />

RESPONSE FACTORS<br />

----------------<br />

SELECTED=0<br />

MODE=FACTOR<br />

RF0 (PID)=DEFAULT 1.00 0.00<br />

RF1 (PID)=---------- 1.00 0.00<br />

RF2 (PID)=---------- 1.00 0.00<br />

RF3 (PID)=---------- 1.00 0.00<br />

RF4 (PID)=---------- 1.00 0.00<br />

RF5 (PID)=---------- 1.00 0.00<br />

RF6 (PID)=---------- 1.00 0.00<br />

RF7 (PID)=---------- 1.00 0.00<br />

RF8 (PID)=---------- 1.00 0.00<br />

RF9 (PID)=---------- 1.00 0.00<br />

RF0 (FID)=DEFAULT 1.00 0.00<br />

RF1 (FID)=---------- 1.00 0.00<br />

RF2 (FID)=---------- 1.00 0.00<br />

RF3 (FID)=---------- 1.00 0.00<br />

RF4 (FID)=---------- 1.00 0.00<br />

RF5 (FID)=---------- 1.00 0.00<br />

RF6 (FID)=---------- 1.00 0.00<br />

RF7 (FID)=---------- 1.00 0.00<br />

RF8 (FID)=---------- 1.00 0.00<br />

RF9 (FID)=---------- 1.00 0.00<br />

ALARMS<br />

------<br />

STEL (PID)= 0.00 PPM<br />

LOW (PID)= 0.00 PPM<br />

HIGH (PID)= 0.00 PPM<br />

STEL (FID)= 0.00 PPM<br />

LOW (FID)= 0.00 PPM<br />

HIGH (FID)= 0.00 PPM<br />

USER OPTIONS<br />

------------<br />

PROBE=BASIC<br />

BARCODE READER=NONE<br />

KEY CLICK=OFF<br />

DISPLAY DELAY=MEDIUM<br />

PC LINK TYPE=TEXT-AUTO<br />

DIGITAL DATA LOGGING=NONE<br />

END<br />

Calibration File Format<br />

The calibration header format is as follows:<br />

Format<br />

CALIBRATION DATA<br />

VER=x.x<br />

<br />

<br />

<br />

<br />

END<br />

Comments<br />

Title, not processed<br />

x=0-9<br />

Blank line, not processed<br />

Blank line, not processed<br />

PID if available, see below<br />

FID if available, see below<br />

Marks end-of-text<br />

The calibration data format is as follows (note: from 1 to 9 span gas points are allowed):<br />

ddd CALIBRATION:<br />

<br />

CALIB KNOWN CALIB MEASURED<br />

TYPE CONC DATE/TIME CALIB VALUE<br />

----- ---------- ------------------ -------------------------<br />

ZERO aaaaaa uuu dd mmm yy hh:mm:ss bbbbbbbb COUNTS fffffffff<br />

SPAN1 aaaaaa uuu dd mmm yy hh:mm:ss bbbbbbbb COUNTS fffffffff<br />

. . . . . . . . . . .<br />

. . . . . . . . . . .<br />

. . . . . . . . . . .<br />

SPANn aaaaaa uuu dd mmm yy hh:mm:ss bbbbbbbb COUNTS fffffffff<br />

<br />

<br />

where:<br />

ddd = detector type<br />

“FID”<br />

“PID”<br />

aaaaaa = gas concentration, floating point, right justified,<br />

leading spaces<br />

uuu = concentration units:<br />

“PPB”, parts-per-billion<br />

84