MI 611-185 – June 1996<br />

Startup and Familiarization<br />

12. Wait to stabilize.<br />

13. Press ENTER = start.<br />

14. Press 4 = Span<br />

15. (PID 1st) Press 2 = PID.<br />

16. Press ENTER = start.<br />

17. Challenge analyzer with isobutylene span gas and wait for readings to stabilize.<br />

18. Press ENTER to accept.<br />

19. Press 4 = Span.<br />

20. Press 3 = FID.<br />

21. Press ENTER = Start.<br />

22. Challenge analyzer with methane span gas and wait for readings to stabilize.<br />

23. Press ENTER = Accept.<br />

24. Press 5 = Response Factor.<br />

25. Confirm that Response Factor says “RF0:DEFAULT”<br />

26. Press EXIT 2 times to main menu.<br />

27. Press 1 = Run.<br />

You are now in the survey mode.<br />

NOTE: To perform more sophisticated operations, you will need to read the rest of the<br />

manual .<br />

To power down this instrument, simply press and hold the OFF key. With FID versions, you<br />

must also shut off the gas valve to avoid depleting the tank supply.<br />