MI 611-185 – June 1996<br />

Introduction<br />

Table 1. Specifications (Continued)<br />

Item<br />

Response Time using<br />

charcoal filter adapter<br />

Recovery Time using<br />

charcoal filter adapter<br />

Data Storage Interval<br />

Sample Flow Rate<br />

Battery<br />

Battery Charger<br />

PID Lamp Life<br />

FID Life<br />

Audio Output Level<br />

Gas Cylinder Capacity<br />

Hydrogen Supply<br />

Operating Time<br />

Enclosure Description<br />

Specification<br />

PID Instrument — Less than 20 seconds for 90% of final value, using<br />

100 ppm of isobutylene<br />

FID Instrument — Less than 20 seconds for 90% of final value, using<br />

10,000 ppm of methane<br />

PID Instrument — Less than 20 seconds to return to 10% of original<br />

value, using 100 ppm of isobutylene<br />

FID Instrument — Less than 20 seconds to return to 10% of original<br />

value, using 10,000 ppm of methane<br />

Auto Mode — 1 per second to 1 per 999 minutes, user selectable<br />

VOC or FE Mode — 2 to 30 seconds, user-selectable<br />

1 liter/minute, nominal, at sample probe inlet<br />

The battery can be fully charged in less than 16 hours. The battery<br />

operating time is 8 hours minimum at 20° (32°F). Use of the<br />

backlight on the probe display shortens battery life. The battery is<br />

replaceable by removing the bottom cover of the instrument.<br />

(WARNING: Do not replace battery in an area classified as hazardous<br />

due to presence of flammable gases or vapors.)<br />

The battery charger is a separate unit capable of operating the<br />

analyzer while simultaneously charging the internal battery. An<br />

adapter cable is provided to charge the battery separate from the<br />

analyzer. The charger can charge a fully discharged battery in a<br />

maximum of 16 hours. Charging takes longer if performed while the<br />

instrument is operating. Two types of chargers are available, one for<br />

120 V ac, 50/60 Hz and another for 230 V ac, 50/60 Hz.<br />

(WARNING: Do not operate battery charger in an area classified as<br />

hazardous due to presence of flammable gases or vapors.)<br />

Greater than 2000 hours for 10.6 eV lamp, with normal cleaning<br />

Greater than 2000 hours<br />

Greater than 75 dB at 3 feet<br />

Pressure — 15.3 MPa at 25°C (2200 psi at 77°F) maximum<br />

Empty — 85 cc (5.19in 3 )<br />

Greater than 8 hours of continuous operation, starting from a<br />

cylinder charged up to 15.3 MPa (2200 psi)<br />

The analyzer enclosure and front panel are made from a chemically<br />

resistant thermoplastic material. The approximate dimensions are<br />

343 x 262 x 81mm (13.5 x 10.3 x 3.2 in). The enclosure is designed<br />

to provide environmental protection.<br />