MI 611-185 – June 1996<br />
Appendix A: Text Data Formats<br />
NOTE: Downloaded route entries may be terminated any time after the tag by the<br />
. Partial text entries are processed as ending in trailing spaces. Partial<br />
number entries are processed as if complete. The default value for fields not included<br />
in the entry are:<br />
Field<br />
Default<br />
Area<br />
all spaces<br />
Subarea<br />
all spaces<br />
Leak rate 0.0<br />
Response factor 1.0<br />
Equipment type all spaces<br />
Equipment size all spaces<br />
Description all spaces<br />
Sample Route List File<br />
ROUTE LIST<br />
VER= 1.00<br />
RUN ID=0001<br />
LEAK RESP EQP EQUIP<br />
TAG AREA SUBAREA RATE FACT TYP SIZE DESCRIPTION<br />
---------------- -------- -------- ----- ---- --- ------ -----------<br />
1 AREA1 SUBAREA1 0 1.00 N/A 1.00 DESCRIPTION 1<br />
2 AREA2 SUBAREA2 0 1.00 BON 0.50 DESCRIPTION 2<br />
3 AREA3 SUBAREA3 0 1.00 CSL 1.75 DESCRIPTION 3<br />
4 AREA4 SUBAREA4 0 1.00 DFL 10.00 DESCRIPTION 4<br />
5 AREA5 SUBAREA5 0 1.00 PKG 5.50 DESCRIPTION 5<br />
6 AREA6 SUBAREA6 0 1.00 PSL 0.25 DESCRIPTION 6<br />
7 AREA7 SUBAREA7 0 1.00 RPL 950.50 DESCRIPTION 7<br />
8 AREA8 SUBAREA8 0 1.00 VLV 0.25 DESCRIPTION 8<br />
9 AREA9 SUBAREA9 0 1.00 FLG 0.75 DESCRIPTION 9<br />
10 AREA10 S-AREA10 0 1.00 FLG 0.75 DESCRIPTION 10<br />
11 AREA11 S-AREA11 0 1.00 VLV 0.75 DESCRIPTION 11<br />
12 AREA12 S-AREA12 0 1.00 PRV 1.50 DESCRIPTION 12<br />
13 AREA13 S-AREA13 0 1.00 VLV 0.75 DESCRIPTION 13<br />
14 AREA14 S-AREA14 0 1.00 CMP 0.00 DESCRIPTION 14<br />
15 AREA15 S-AREA15 0 2.00 VLV 0.75 DESCRIPTION 15<br />
16 AREA16 S-AREA16 0 5.00 VLV 0.75 DESCRIPTION 16<br />
17 AREA17 S-AREA17 0 .001 VLV 1.50 DESCRIPTION 17<br />
18 AREA18 S-AREA18 0 .01 FLG 1.50 DESCRIPTION 18<br />
19 AREA19 S-AREA19 0 0.01 FLG 0.75 DESCRIPTION 19<br />
20 AREA20 S-AREA20 0 .1 VLV 0.75 DESCRIPTION 20<br />
21 AREA21 S-AREA21 5000 00.1 VLV 0.75 DESCRIPTION 21<br />
22 AREA22 S-AREA22 99999 1 VLV 0.75 DESCRIPTION 22<br />
23 AREA23 S-AREA23 10000 10 VLV 0.75 DESCRIPTION 23<br />
24 AREA24 S-AREA24 40000 100 FLG 0.50 DESCRIPTION 24<br />
25 AREA25 S-AREA25 50 1000 VLV 0.75 DESCRIPTION 25<br />
END<br />
Configuration File Format<br />
The configuration format is as follows (options are separated by a '|'):<br />
Format<br />
CONFIGURATION<br />
VER=X.X<br />
<br />
<br />
MODEL=xxxxxxxxxx<br />
SERIAL NUMBER=ididididid<br />
DETECTOR=PID|FID|PID&FID<br />
OPERATOR ID=opopopopop<br />
<br />
LOGGING<br />
-------<br />
LOG MODE=NONE|AUTO|VOC|FE|CUSTOM<br />
LOG AUTO REPEAT=OFF|ON<br />
LOG UNIT LOCK=OFF|PPB|PPM|PERCENT<br />
LOG SAVE MODE=MANUAL|AUTO<br />
LOG STORAGE FORMAT=NONE|FOXAUTO|FOXVOC|FEVOC<br />
LOG SAMPLE TIME=nnnnn<br />
82