Display Menus MI 611-185 – June 1996<br />
In the LOG AUTO mode, data is logged in the format described below:<br />
1. Every time RUN mode is entered a new header is created.<br />
2. Data is logged periodically, at the user-entered log rate, in the following format:<br />
Date Time PID (PPM)Alarms FID (PPM) Alarms<br />
7/28/93 11:20:57:00 187.1 L 157.1 L<br />
7/28/93 11:21:29:00 436.6 H 519.6 L<br />
7/28/93 11:21:54:00 708.4 LH 847.0 LH<br />
7/28/93 11:22:19:00 300.7 L 381.9 L<br />
7/28/93 11:22:36:00 223.8 265.1<br />
Log VOC<br />
Approximately 9500 data samples may be taken with a single header.<br />
1. From the LOG MODE Menu, press 3=VOC/FE, then 1=VOC.<br />
LOG MODE<br />
Auto,rate=12 sec<br />
1=None 2=Auto<br />
3=VOC 4=FE<br />
Press 3, then 1 to display this message.<br />
Enter Sampling:<br />
Voc, Time=00s, Max<br />
Up/Dn=Next type<br />
Enter=ACCEPT<br />
2. Press the Up/Dn arrow keys to choose the type of sampling to be used (Max, Avg,<br />
Smp).<br />
a. Max=the maximum reading obtained during the log interval.<br />
b. Avg=the average reading obtained during the log interval.<br />
c. Smp=the last reading obtained during the log interval.<br />
3. Type the appropriate time interval desired for VOC sampling.<br />
4. Press ENTER, and, after a brief message of acceptance, the VOC method of logging<br />
is now ready to be used in the RUN Mode.<br />
NOTE: The allowable Sample Time range is 2-30 seconds. If a Sample Time of less<br />
than 2 seconds or greater than 30 seconds is selected, a warning message will appear<br />
and you will be required to re-enter the Sample Time.<br />
41