MI 611-185 – June 1996<br />
Display Menus<br />
Calibration Configuration<br />
Before you calibrate the <strong>TVA</strong>-<strong>1000B</strong> for the first time, you may want to customize certain calibration<br />
settings. Once you have configured these settings, you don’t need to set them again<br />
every time you calibrate unless you want to change one.<br />
In order to set your calibration options from the CALIBRATION menu, choose menu selection<br />
#1 (Cfg). This will produce the following CAL CONFIG MENUs which can be scrolled<br />
through by using the Up and Down arrows:<br />
Cal Config MENU<br />
1= Number Span Pt.<br />
2=Backgnd corrct<br />
Up/Dn=More<br />
Cal Config MENU<br />
1= Accept Mode<br />
2=Save Mode<br />
Up/Dn=More<br />
Cal Config MENU<br />
1=RF calc mode<br />
Up/Dn=More<br />
Number Span Pt<br />
This selection allows you to determine the number of span points that will be used to calibrate<br />
your <strong>TVA</strong>-<strong>1000B</strong>. Choosing this selection will produce the following screen:<br />
PID: 1<br />
FID: 1<br />
Span Pts. 1=Both<br />
2=PID 3=FID<br />
From this screen, you may choose which detector you wish to set the number of span points<br />
for. Choosing one of these selections produces the following screens:<br />
FID only PID only FID/PID<br />
Number Span Pts:<br />
FID: 1<br />
Select: 1 to 9<br />
Number Span Pts:<br />
PID: 1<br />
Select: 1 to 9<br />
Number Span Pts:<br />
P&F: 1<br />
Select: 1 to 9<br />
From these screens, press the number of span points that you wish to use. Both detectors can<br />
have up to 9 span points each, and they do not necessarily need to have the same number of<br />
span points.<br />
IMPORTANT NOTE: If you choose to use more than one span point, data logged by<br />
the <strong>TVA</strong>-<strong>1000B</strong> cannot be downloaded to the original PC software supplied with old<br />
style <strong>TVA</strong>-1000A units. You must use the new PCIP software supplied with your<br />
new <strong>TVA</strong>-<strong>1000B</strong> to download this data.<br />
26