MI 611-185 – June 1996<br />

Display Menus<br />

Calibration Configuration<br />

Before you calibrate the <strong>TVA</strong>-<strong>1000B</strong> for the first time, you may want to customize certain calibration<br />

settings. Once you have configured these settings, you don’t need to set them again<br />

every time you calibrate unless you want to change one.<br />

In order to set your calibration options from the CALIBRATION menu, choose menu selection<br />

#1 (Cfg). This will produce the following CAL CONFIG MENUs which can be scrolled<br />

through by using the Up and Down arrows:<br />

Cal Config MENU<br />

1= Number Span Pt.<br />

2=Backgnd corrct<br />

Up/Dn=More<br />

Cal Config MENU<br />

1= Accept Mode<br />

2=Save Mode<br />

Up/Dn=More<br />

Cal Config MENU<br />

1=RF calc mode<br />

Up/Dn=More<br />

Number Span Pt<br />

This selection allows you to determine the number of span points that will be used to calibrate<br />

your <strong>TVA</strong>-<strong>1000B</strong>. Choosing this selection will produce the following screen:<br />

PID: 1<br />

FID: 1<br />

Span Pts. 1=Both<br />

2=PID 3=FID<br />

From this screen, you may choose which detector you wish to set the number of span points<br />

for. Choosing one of these selections produces the following screens:<br />

FID only PID only FID/PID<br />

Number Span Pts:<br />

FID: 1<br />

Select: 1 to 9<br />

Number Span Pts:<br />

PID: 1<br />

Select: 1 to 9<br />

Number Span Pts:<br />

P&F: 1<br />

Select: 1 to 9<br />

From these screens, press the number of span points that you wish to use. Both detectors can<br />

have up to 9 span points each, and they do not necessarily need to have the same number of<br />

span points.<br />

IMPORTANT NOTE: If you choose to use more than one span point, data logged by<br />

the <strong>TVA</strong>-<strong>1000B</strong> cannot be downloaded to the original PC software supplied with old<br />

style <strong>TVA</strong>-1000A units. You must use the new PCIP software supplied with your<br />

new <strong>TVA</strong>-<strong>1000B</strong> to download this data.<br />

26