WATERTRAP Probe MI 611-182 – March 1996<br />

Membrane Life<br />

The life of the WATERTRAP membrane depends largely upon the conditions the membrane<br />

is exposed to. Theoretically, the membrane can last several months under normal ambient<br />

conditions; however, this lifetime will be shortened considerably by prolonged exposure to liquids,<br />

particulates, high temperatures, and high humidity.<br />

Membrane Replacement Suggestions<br />

It is recommended to verify the integrity of the membrane each time your analyzer is used. It<br />

is also recommended to replace the membrane each month if the OVA or <strong>TVA</strong>-1000 is used<br />

daily. Otherwise, the membrane should be replaced during quarterly instrument evaluations.<br />

The membrane should also be replaced after contact with liquid. In some instances, a membrane<br />

can be restored after contact with only water by shaking the water from the probe.<br />

However, if the membrane is contaminated by an oily or high surface tension liquid, it should<br />

be replaced immediately and the entire probe assembly thoroughly cleaned. Under dusty conditions,<br />

membrane operation should be checked frequently. Finally, the membrane should be<br />

replaced if it becomes torn, ruptured, or if its surface becomes coated by foreign material.<br />

3