Troubleshooting MI 611-185 – June 1996<br />
Problem Possible Reason Solution/Action<br />
PID lamp not operating<br />
Bad PID lamp<br />
Replace PID capsule.<br />
PID noisy<br />
Bad PID lamp<br />
Replace PID capsule.<br />
Unable to calibrate<br />
PID<br />
Slow response time<br />
High background<br />
readings<br />
Dirty PID window<br />
Erratic pump flow<br />
Bad calibration<br />
Pump not on<br />
PID lamp not on<br />
Span concentration not properly<br />
set<br />
Cal gases contaminated<br />
Sample line/filter cups contaminated<br />
PID window dirty or capsule<br />
contaminated or faulty<br />
Internal detector fault or contamination<br />
Insufficient sample flow<br />
Sampling non-volatile compounds<br />
High ambient concentration<br />
Zero drift/improper calibration<br />
Sample line contamination<br />
Detector capsule contamination<br />
Clean PID window.<br />
Clean/replace filter cups.<br />
Clean/replace flame arrestor.<br />
Ensure proper calibration.<br />
Turn pump on.<br />
Turn PID lamp on.<br />
Input correct span gas<br />
concentration at CAL menu.<br />
Use clean cal gases and sampling equipment.<br />
Clean/replace sample line.<br />
Clean/replace filter cups.<br />
Clean PID window.<br />
Replace PID capsule.<br />
Contact TEI authorized service center.<br />
Clean/replace filter cups.<br />
Clean/replace flame arrestor.<br />
Clean/replace sample line.<br />
Contact TEI Applications Laboratory.<br />
N/A.<br />
Ensure proper zero/span calibration.<br />
Clean/replace sample line.<br />
Clean/replace filter cups.<br />
Clean/replace FID capsule.<br />
Clean PID window.<br />
Replace PID capsule.<br />
75