MI 611-185 – June 1996<br />

Introduction<br />

l<br />

External<br />

Influence<br />

Ambient<br />

Temperature<br />

Ambient<br />

Pressure<br />

Relative<br />

Humidity<br />

Radiated<br />

Susceptibility<br />

Conducted<br />

Susceptibility<br />

Conducted<br />

Emission<br />

Radiated<br />

Emission<br />

ESD Sensitivity<br />

Battery Charger<br />

Supply Voltage<br />

Battery Charger<br />

Supply<br />

Frequency<br />

Reference<br />

Operating<br />

Conditions<br />

23±2 °C<br />

73±2 °F<br />

860 to<br />

1060 mbar<br />

Table 2. External Influences<br />

Normal<br />

Operating<br />

Conditions<br />

0 to +40°C<br />

32 to 104 °F<br />

50%±10% FID: 20 to 95%<br />

PID: 20 to 70%<br />

noncondensing<br />

None<br />

120 or 230 ±1%<br />

Vac<br />

50/60 Hz<br />

±0.5 Hz<br />

Operative<br />

Limits<br />

0 and +50°C<br />

32 and 122 °F<br />

Transportation<br />

and Storage<br />

Limits<br />

-20 to +60°C<br />

-4 and 140 °F<br />

70 to 108 kPa 20 to 108 kPa<br />

15 and 95%<br />

noncondensing<br />

30 V/m<br />

27 to 500 MHz<br />

Not Applicable, Battery Operated<br />

Not Applicable, Battery Operated<br />

0 to 100%<br />

80 dBµV<br />

0.15 to 30 MHz<br />

>6000 Volts<br />

120 or 230<br />

Not Applicable<br />

+15%, -10%<br />

Vac<br />

47 to 63 Hz 47 and 63 Hz Not Applicable<br />

Product Safety Specifications<br />

Table 3. Product Safety Specifications<br />

Testing Laboratory, Types of Protection,<br />

and Area Classification<br />

FM: intrinsically safe for Class I, Division 1,<br />

Groups A, B, C, and D<br />

*CENELEC: intrinsically safe for Zone 1 and<br />

Zone 2, (EEx ib d IIC)<br />

*CSA: intrinsically safe for Class I, Division 1,<br />

Groups A, B, C, and D.<br />

Conditions of<br />

Classification<br />

Temperature Class T4<br />

Temperature Class T4<br />

Temperature Class T4<br />

*Not available at release. Contact <strong>Thermo</strong> Environmental Instruments for current status.<br />

Electrical<br />

Classification<br />

Code<br />

FAZ<br />

EBZ<br />

CAZ<br />

6