MI 611-185 – June 1996<br />

Hardware<br />

A segmented analog bargraph that represents a logarithmic scale for the total analyzer range<br />

appears below the digital display. This display, which may be backlit under low light conditions, is<br />

active only in the RUN mode. In all other modes, it displays OFF.<br />

The measured value display area is used to flash logging prompts during survey and log mode. In<br />

addition, errors on the sidepack are indicated by “Err” on the probe. Errors are cleared by using<br />

the LOG key.<br />

MEASURED VALUE AND ERROR/LOGGING PROMPTS<br />

MEASUREMENT UNITS<br />

DETECTOR<br />

BEING<br />

DISPLAYED<br />

8.8.8.8.<br />

ppm<br />

% ppb<br />

fid<br />

pid<br />

>>><br />

OVERRANGE INDICATOR<br />

LOGARITHMIC<br />

BARGRAPH<br />

DET<br />

LOG<br />

DETECTOR SELECTOR<br />

BUTTON<br />

LOG SELECTOR<br />

BUTTON<br />

BUTTON FOR<br />

DISPLAY BACKLIGHTING<br />

Figure 8. <strong>TVA</strong>-<strong>1000B</strong> BASIC Probe Display<br />

NOTE: Use of the backlight draws additional power from the battery and will<br />

shorten the runtime of the <strong>TVA</strong>-<strong>1000B</strong>.<br />

BASIC Probe Keys<br />

The BASIC probe keypad has three keys. The keys are labeled with DET, LOG, and a lamp symbol.<br />

Key<br />

Function<br />

DET Press and release this key to select the display of either the FID or the PID<br />

readings in the RUN mode.<br />

Press and hold this key (approx. 2 seconds) to cancel selected logging modes.<br />

LOG Press and release this key to start the datalogging feature. Survey information is<br />

then automatically stored in the analyzer memory. The selection in SETUP/Log<br />

must be AUTO or VOC.<br />

Press and release this key to clear a sidepack error when the display reads “ Err”.<br />

Lamp Symbol Press this key to turn the backlight for the probe display on or off. You should use<br />

it only during low light conditions as it draws additional power from the<br />

battery and thus shortens instrument run time.<br />