MI 611-185 – June 1996<br />

Accessories<br />

To replace charcoal in a charcoal filter, use the following procedure:<br />

1. Remove the sampling assembly from the probe by loosening the probe nut and<br />

pulling the sampling assembly free from the probe.<br />

2. Unscrew the cap from the sampling assembly. This will expose the end of the charcoal<br />

filter cartridge. See diagram below.<br />

3. Tip the cartridge up and empty the charcoal from the cartridge.<br />

4. Clean the inside of the cartridge with a cotton swab dipped in isopropyl alcohol.<br />

5. Dry the cartridge thoroughly.<br />

6. Re-fill the cartridge with new “dust-free” charcoal of the type specified for the filter.<br />

7. Apply Teflon tape to the male threads of the charcoal cartridge and screw on the<br />

cap.<br />

8. Re-insert the sampling assembly into the probe and tighten the probe nut.<br />

CARTRIDGE<br />

FILLED WITH<br />

ACTIVATED<br />

CHARCOAL<br />

THREADED<br />

END WITH<br />

TEFLON<br />

TAPE<br />

CAP<br />

Figure 14. Activated Charcoal Filter Adapter<br />

For information on additional accessories, such as calibration kits, enhanced probes, barcode<br />

readers, optional energy lamps (PID), or dilutor kits, please contact TEI.<br />

80