MI 611-185 – June 1996<br />
Accessories<br />
To replace charcoal in a charcoal filter, use the following procedure:<br />
1. Remove the sampling assembly from the probe by loosening the probe nut and<br />
pulling the sampling assembly free from the probe.<br />
2. Unscrew the cap from the sampling assembly. This will expose the end of the charcoal<br />
filter cartridge. See diagram below.<br />
3. Tip the cartridge up and empty the charcoal from the cartridge.<br />
4. Clean the inside of the cartridge with a cotton swab dipped in isopropyl alcohol.<br />
5. Dry the cartridge thoroughly.<br />
6. Re-fill the cartridge with new “dust-free” charcoal of the type specified for the filter.<br />
7. Apply Teflon tape to the male threads of the charcoal cartridge and screw on the<br />
cap.<br />
8. Re-insert the sampling assembly into the probe and tighten the probe nut.<br />
CARTRIDGE<br />
FILLED WITH<br />
ACTIVATED<br />
CHARCOAL<br />
THREADED<br />
END WITH<br />
TEFLON<br />
TAPE<br />
CAP<br />
Figure 14. Activated Charcoal Filter Adapter<br />
For information on additional accessories, such as calibration kits, enhanced probes, barcode<br />
readers, optional energy lamps (PID), or dilutor kits, please contact TEI.<br />
80