MI 611-182 – March 1996<br />
WATERTRAP Probe<br />
WATERTRAP Performance<br />
The WATERTRAP probe will allow vapors (including water vapors) to pass through to the<br />
analyzer unhindered while blocking the passage of water and other more viscous liquids. The<br />
membrane will not reduce flow to the OVA or <strong>TVA</strong>-1000 (unless the membrane becomes<br />
coated or contaminated), and it will not be corroded by most chemical vapors.<br />
NOTE: The WATERTRAP membrane is not chemically compatible with hydrofluoric<br />
acid.<br />
Membrane Installation <strong>Instruction</strong>s<br />
NOTE: Take care not to touch or damage the membrane surface.<br />
1. Loosen the probe nut and remove the WATERTRAP probe from the OVA or<br />
<strong>TVA</strong>-1000 readout assembly.<br />
2. Disassemble the WATERTRAP probe by grasping the probe’s lower half and turning<br />
the probe half counterclockwise with probe tip in upright position.<br />
3. Insert the tip of the probe into the back end of the probe’s lower half.<br />
4. Use the probe tip to push the membrane support and membrane/O-ring assembly<br />
gently out of the lower half of the probe body.<br />
5. Remove and discard the used membrane/O-ring assembly.<br />
6. Re-install the membrane support into the lower half of the probe body with the<br />
rough side facing up.<br />
NOTE: Take special care not to contact or damage the membrane during reassembly.<br />
7. Place a new membrane/O-ring assembly over the membrane support and align the<br />
O-ring over the O-ring groove. The O-ring should be at the top of the membrane.<br />
8. Carefully press the O-ring into the groove at two opposite locations.<br />
9. Repeat the previous step at different locations until the entire O-ring is depressed<br />
approximately one half of its depth. The membrane should not be tightly<br />
stretched.<br />
10. Reassemble the probe by grasping each end by hand and turning clockwise until<br />
tight.<br />
11. Reinstall the WATERTRAP probe in the OVA or <strong>TVA</strong>-1000 and tighten the<br />
probe nut.<br />
2