MI 611-182 – March 1996<br />

WATERTRAP Probe<br />

WATERTRAP Performance<br />

The WATERTRAP probe will allow vapors (including water vapors) to pass through to the<br />

analyzer unhindered while blocking the passage of water and other more viscous liquids. The<br />

membrane will not reduce flow to the OVA or <strong>TVA</strong>-1000 (unless the membrane becomes<br />

coated or contaminated), and it will not be corroded by most chemical vapors.<br />

NOTE: The WATERTRAP membrane is not chemically compatible with hydrofluoric<br />

acid.<br />

Membrane Installation <strong>Instruction</strong>s<br />

NOTE: Take care not to touch or damage the membrane surface.<br />

1. Loosen the probe nut and remove the WATERTRAP probe from the OVA or<br />

<strong>TVA</strong>-1000 readout assembly.<br />

2. Disassemble the WATERTRAP probe by grasping the probe’s lower half and turning<br />

the probe half counterclockwise with probe tip in upright position.<br />

3. Insert the tip of the probe into the back end of the probe’s lower half.<br />

4. Use the probe tip to push the membrane support and membrane/O-ring assembly<br />

gently out of the lower half of the probe body.<br />

5. Remove and discard the used membrane/O-ring assembly.<br />

6. Re-install the membrane support into the lower half of the probe body with the<br />

rough side facing up.<br />

NOTE: Take special care not to contact or damage the membrane during reassembly.<br />

7. Place a new membrane/O-ring assembly over the membrane support and align the<br />

O-ring over the O-ring groove. The O-ring should be at the top of the membrane.<br />

8. Carefully press the O-ring into the groove at two opposite locations.<br />

9. Repeat the previous step at different locations until the entire O-ring is depressed<br />

approximately one half of its depth. The membrane should not be tightly<br />

stretched.<br />

10. Reassemble the probe by grasping each end by hand and turning clockwise until<br />

tight.<br />

11. Reinstall the WATERTRAP probe in the OVA or <strong>TVA</strong>-1000 and tighten the<br />

probe nut.<br />

2