MI 611-185 – June 1996<br />
Display Menus<br />
The ACCEPTED message appears for a short time and is then replaced by the<br />
normal CALIBRATION menu.<br />
When the ACCEPTED message disappears and the CALIBRATION menu<br />
appears, the ZERO reference value is stored. This value is stored in non-volatile<br />
memory until the next calibration is performed. The date and time of this calibration<br />
are stored and can be accessed through the INFO menu.<br />
NOTE: For optimum accuracy, re-zero the FID every time the hydrogen supply valve<br />
is turned on.<br />
Span Reference Point(s)<br />
To set the span reference point, execute the procedure described below. Note that the<br />
procedure is the same as that for setting the zero reference except that a span gas is<br />
used instead of a zero gas. The procedure is:<br />
1. From the CALIBRATION MENU display, press 4=Span.<br />
NOTES:<br />
1. The following sequence shows the procedure when the <strong>TVA</strong>-<strong>1000B</strong> is configured<br />
with both Cal mode and Cal Save mode = Auto. If either mode is <strong>Manual</strong>, an extra<br />
confirmation is required at the appropriate step.<br />
2. If your instrument is a dual detector type (PID and FID), you can set the span reference<br />
for the PID and FID separately, both together, or one in one type of calibration<br />
gas and one in another type of gas. To do so, follow the same procedure.<br />
3. If multiple span points are used, repeat the following procedure for each and every<br />
span point.<br />
When you press 4=Span, the display will show:<br />
FID-only PID-only PID/FID<br />
FID 7654321<br />
Span P+1 (Up/Dn)<br />
Enter=New Calib<br />
PID 7654321<br />
Span P+1 (Up/Dn)<br />
Enter=New Calib<br />
PID 7654321<br />
FID 7654321<br />
Span Pt1 (Up/Dn)<br />
1==Both 2Pid 3Fid<br />
If more than one reference span gas concentration has been configured, pressing<br />
the Up/Dn keys will allow you to scroll through all defined reference gas concentration<br />
points.<br />
34