Display Menus MI 611-185 – June 1996<br />
2. To change the alarm level to a new value, press ENTER on a single detector instrument<br />
or 1, 2, or 3 on a dual detector instrument and see:<br />
FID-only PID-only PID/FID<br />
Enter STEL:<br />
FID: 000.00 ppm<br />
Up/Dn=Next Unit<br />
Enter=Accept<br />
Enter STEL:<br />
PID:000.00 ppm<br />
Up/Dn=Next Unit<br />
Enter=Accept<br />
Enter STEL:<br />
P&F:000.00 ppm<br />
Up/Dn=Next Unit<br />
Enter=Accept<br />
Use the up and down arrow keys to select %, PPM, PPB, and decimal point position, then<br />
type the numeric value for the alarm level desired.<br />
3. Press ENTER to store new values into instrument memory.<br />
4. Press EXIT to return to the ALARM LEVELS without making a change.<br />
Low Ceiling Alarm<br />
1. From the ALARM LEVELS display, press 2=Low Ceiling. The previous alarm settings<br />
for the Low Ceiling alarm are then displayed, as follows.<br />
FID-only PID-only PID/FID<br />
FID: 80.00 ppm<br />
Low Ceiling Alarm<br />
Enter=New Value<br />
PID: 80.00 ppm<br />
Low Ceiling Alarm<br />
Enter=New Value<br />
PID: 80.00 ppm<br />
FID: 80.00 ppm<br />
Low Ceiling: 1=Both<br />
2=PID 3=FID<br />
2. To change the alarm level to a new value on a single detector instrument, press<br />
ENTER; on a dual detector instrument, press 1, 2, or 3, and see:<br />
Enter low ceil:<br />
FID: 000.00 ppm<br />
Up/Dn=Next Unit<br />
Enter=Accept<br />
Enter low ceil:<br />
PID:000.00 ppm<br />
Up/Dn=Next Unit<br />
Enter=Accept<br />
Enter low ceil:<br />
P&F:000.00 ppm<br />
Up/Dn=Next Unit<br />
Enter=Accept<br />
Use the up and down arrow keys to select %, PPM, PPB, and decimal point position,<br />
and then type the numeric value for the alarm level desired.<br />
3. Press ENTER to store the new values into instrument memory.<br />
4. Press EXIT to return to the ALARM LEVELS.<br />