20.11.2014 Views

Thermo Scientific TVA-1000B Instruction Manual - Geotech ...

Thermo Scientific TVA-1000B Instruction Manual - Geotech ...

Thermo Scientific TVA-1000B Instruction Manual - Geotech ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

MI 611-185 – June 1996<br />

Display Menus<br />

High Ceiling Alarm<br />

1. From the ALARM LEVELS display, press 3=High Ceiling. The previous alarm<br />

settings for the High Ceiling alarm are displayed.<br />

FID-only PID-only PID/FID<br />

FID: 80.00 ppm<br />

High Ceiling Alarm<br />

Enter=New Value<br />

PID: 80.00 ppm<br />

High Ceiling Alarm<br />

Enter=New Value<br />

PID: 80.00 ppm<br />

FID: 80.00 ppm<br />

High Ceiling: 1=Both<br />

2=PID 3=FID<br />

2. To change the alarm level to a new value, press ENTER on a single detector<br />

instrument or 1, 2, or 3 on a dual detector instrument and see:<br />

Enter high ceil:<br />

FID: 000.00 ppm<br />

Up/Dn=Next Unit<br />

Enter=Accept<br />

Enter high ceil:<br />

PID:000.00 ppm<br />

Up/Dn=Next Unit<br />

Enter=Accept<br />

Enter high ceil:<br />

P&F:000.00 ppm<br />

Up/Dn=Next Unit<br />

Enter=Accept<br />

Use the up and down arrow keys to select %, PPM, PPB, and decimal point position,<br />

and then type the numeric value for the alarm level desired.<br />

3. Press ENTER to store the new values into instrument memory.<br />

4. Press EXIT to return to the ALARM LEVELS display.<br />

Log Methods<br />

Selecting the LOG mode not only identifies the way in which data is stored in instrument<br />

memory but also governs the way the RUN mode works. Four possible choices in the LOG<br />

mode are available. Each is explained below.<br />

Option<br />

Description<br />

1. NONE Selecting None means that no logging action takes place at all. When the<br />

instrument is in the RUN mode, you can only display readings.<br />

2. AUTO When you select and configure the AUTO method, the RUN mode functions<br />

automatically once an action is initiated. You choose an automatic logging rate<br />

between 1 second and 999 minutes, type a tag* identifier, and start the instrument.<br />

You initiate automatic sampling by either pressing the LOG key on the<br />

probe or the ENTER key on the instrument. From that point on, the instrument<br />

automatically logs the reading of the sample at the end of each countdown<br />

and continues to do so until the battery goes low or you press the EXIT<br />

key.<br />

38

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!