20.11.2014 Views

Thermo Scientific TVA-1000B Instruction Manual - Geotech ...

Thermo Scientific TVA-1000B Instruction Manual - Geotech ...

Thermo Scientific TVA-1000B Instruction Manual - Geotech ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Display Menus MI 611-185 – June 1996<br />

Example: A <strong>TVA</strong>-<strong>1000B</strong> FID is calibrated with zero air and a 100 ppm Methane in air span<br />

gas. The counts observed for the zero are 2895 and the counts observed for the span are<br />

27395. The span sensitivity is thus 245 counts/ppm [(27395-2750)/100 ppm]. Since both of<br />

these values (2895 zero counts and 245 detector sensitivity) are within the acceptable range,<br />

the calibration is a good calibration. Examples of a bad calibration include unusually high<br />

zero counts, or unusually low detector sensitivity. These problems can often be attributed to<br />

poor calibration gases, contaminated sampling accessories, a faulty detector capsule, or failure<br />

to follow the proper calibration procedure. For more information, consult the “Troubleshooting”<br />

guide in this manual or contact TEI for assistance.<br />

Defining the Span Gas Concentration(s)<br />

NOTES:<br />

1. The span gas concentration is the known concentration of the gas standards used to<br />

calibrate your <strong>TVA</strong>. Methane in air is the recommended calibration standard for the<br />

FID, and Isobutylene in air is the recommended calibration standard for the PID.<br />

Other gases may be used if desired.<br />

2. If your instrument is equipped with dual detectors, you may choose to calibrate the<br />

PID and FID separately or together.<br />

3. If your instrument is configured for multiple span points, be sure to set the concentration<br />

for ALL span points.<br />

1. From the CALIBRATION menu display, press 2=SpanConc. The upper display<br />

(or two displays if the unit is a dual detector version) will display the concentration<br />

value of your span gas (expressed as ppb, ppm, or %) as of the last calibration:<br />

FID only PID only FID/PID<br />

FID: 100 ppm<br />

Span Gas Concent<br />

Enter=New conc<br />

PID: 100 ppm<br />

PID: 100 ppm<br />

FID: 100 ppm<br />

Span Gas Concent<br />

Enter=New conc Span Conc 1=Both<br />

2=PID 3=FID<br />

If the <strong>TVA</strong>-<strong>1000B</strong> is configured for multiple calibration points, the span gas concentration<br />

values for Point #1 will be displayed. The Up and Down arrow keys can be used to scroll<br />

through the span gas concentration values for other points:<br />

FID only PID only FID/PID<br />

FID: 100 ppm<br />

Span Pt 1 (Up/Dn)<br />

Enter=New conc<br />

PID: 100 ppm<br />

Span Pt 1 (Up/Dn)<br />

Enter=New conc<br />

PID: 100 ppm<br />

FID: 100 ppm<br />

Span Pt 1 (Up/Dn)<br />

1=Both 2PID 3FID<br />

29

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!