20.11.2014 Views

Thermo Scientific TVA-1000B Instruction Manual - Geotech ...

Thermo Scientific TVA-1000B Instruction Manual - Geotech ...

Thermo Scientific TVA-1000B Instruction Manual - Geotech ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Maintenance MI 611-185 – June 1996<br />

8. Turn the end cap over so the red outer surface lies flat and the gray surface faces<br />

up.<br />

9. Place the new flame arrestor in the center hole.<br />

10. Place a screwdriver on the newly installed flame arrestor.<br />

11. Strike the end of the screwdriver with a hammer to secure the new flame arrestor<br />

in place.<br />

12. Replace the spring.<br />

13. Replace the detector cap on the unit.<br />

Cleaning the FID or PID Detector Cavities<br />

1. Close the hydrogen supply valve on the side of the instrument. Turn the instrument<br />

off.<br />

2. Using the special spanner wrench provided with the tool kit, unscrew the cap<br />

holding the respective detector cap.<br />

3. Using the special extractor tool provided with the tool kit, screw the extractor into<br />

the cartridge.<br />

4. Remove the cartridge by pulling on the extractor. Unscrew the extractor from the<br />

cartridge.<br />

5. Carefully clean the inside of the detector cavity using a cotton swab and isopropyl<br />

alcohol. Be sure to clean the high voltage contacts along the side of the cavity. Be<br />

especially careful around the detector signal collector probe at the rear of the cavity<br />

(and the thermocouple probe in the FID).<br />

6. Dry the inside of the cavity using a low heat gun.<br />

7. Insert the cartridges into their respective cavities by reversing the procedure. Note<br />

that the cartridges must be rotated to properly locate the key tabs.<br />

CAUTION: Do not intermix the detector cartridges.<br />

Cleaning or Replacing a Sintered Metal Filter<br />

To remove the sintered metal filter cup from the close area sampling assembly, simply unscrew<br />

the cap from the sampler and tip the assembly so that the filter falls out. The same filter cup<br />

and spring are also located behind the sample line quick connect at the sidepack assembly. To<br />

remove, unscrew the quick connect so that the filter falls out. The internal spring will not fall<br />

out. Refer to the assembly diagram below.<br />

You may clean the filter by immersing it in isopropyl alcohol or equivalent solvent and/or<br />

swabbing the surface with a Q-tip or cotton swab. After thoroughly cleaning the surface of the<br />

filter, place the filter in an oven and dry it at 40 °C to 55 °C for one hour. You can then reinsert<br />

the filter into the sampling assembly (with the closed end of the filter in first) and screw<br />

in the cap.<br />

69

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!