Display Menus MI 611-185 – June 1996<br />

The following variable settings can be configured when designing a Custom log method:<br />

1. Log data type<br />

This setting allows you to choose the format in which data will be saved. The choices<br />

are as follows:<br />

a. None.<br />

b. Auto – Data stored in a fashion similar to Auto logging for download to the TEI<br />

datalogging software only (tag, calibration information, Detector Counts, time/<br />

date, alarms).<br />

c. VOC – Data stored in a fashion similar to VOC logging for download to the<br />

TEI datalogging software only (Tag, Calibration Information, Detector Counts,<br />

time/date, alarms).<br />

d. FE – Data stored in a fashion similar to FE logging for download to FEMS or<br />

via the Text protocol.<br />

2. Sample time<br />

This setting allows you to adjust the sampling interval.<br />

3. Log time units<br />

This setting allows you to choose the units for the sample interval (seconds or minutes).<br />

4. Log sample type<br />

This setting allows you to choose the way in which sample readings are collected over<br />

the sample rate or interval. The choices are as follows:<br />

a. Last sample — the last reading obtained within the sample interval is stored<br />

b. Max — the maximum reading obtained within the sample interval is stored<br />

c. Avg — the average reading obtained over the sample interval is stored<br />

5. Log unit lock<br />

This setting allows you to decide whether to enable auto-ranging or lock the<br />

<strong>TVA</strong>-1000 units to one of three choices: ppb, ppm, or %.<br />

6. Log auto repeat<br />

This setting allows you to determine whether the <strong>TVA</strong>-<strong>1000B</strong> should automatically<br />

cycle through logging events (auto repeat ON) or pause for a user command before<br />

logging (auto repeat OFF).<br />

7. Log save accept<br />

This setting determines whether the <strong>TVA</strong>-<strong>1000B</strong> prompts the user before saving a<br />

logged reading (<strong>Manual</strong>) or automatically saves the reading without a user prompt<br />

(Auto).<br />

8. Tag active<br />

This setting determines whether or not you want to assign tags to logged readings. If<br />

you wish to assign tags, choose ON. If you wish to skip tagging, choose OFF.<br />

43