Maintenance MI 611-185 – June 1996<br />

Hydrogen Gas Tank<br />

FID instruments are supplied with an 85cc hydrogen gas tank. This tank, which may be pressurized<br />

to 2200 psi maximum at 25°C, will provide 8 hours operation when fully charged.<br />

The tank has an integrally mounted high pressure gauge that can be easily read when the tank<br />

is in or out of the instrument. Install the tank in the instrument by inserting it into the receptacle<br />

on the left side and tightening (left hand threads, tighten counterclockwise) until the<br />

rubber tank boot is flush with the instrument sidepack and a slight resistance is felt. Do not<br />

overtighten.<br />

Fuel Refilling Procedure<br />

WARNING: A safe refill operation means there are no hydrogen leaks. Before any<br />

valves are opened, use a wrench to firmly tighten connections to the hydrogen supply<br />

tanks and the tank fill adapter. If escaping hydrogen is heard during the filling operation,<br />

close all valves and correct the leak before proceeding. Leak test with soapy water<br />

or equivalent.<br />

NOTES:<br />

1. Use prepurified ZERO grade hydrogen (certified total hydrocarbons as methane<br />

Hooray! Your file is uploaded and ready to be published.