Display Menus<br />
In the LOG VOC mode, data is logged in the format described below:<br />
1. Every time RUN mode is entered a new header is created.<br />
2. Data is logged in the following format whenever you press the LOG button:<br />
Date Time PID (PPM) Alarms FID (PPM) Alarms Tag<br />
7/28/93 11:20:57:00 187.1 L 157.1 L TAG1<br />
7/28/93 11:21:29:00 436.6 H 519.6 L TAG2<br />
7/28/93 11:21:54:00 708.4 LH 847.0 LH TAG3<br />
7/28/93 11:22:19:00 300.7 L 381.9 L TAG4<br />
7/28/93 11:22:36:00 223.8 265.1 TAG5<br />
Approximately 4500 data samples may be taken.<br />
Log F.E.<br />
1. From the LOG MODE menu, press 3=VOC/FE, then 2=FE The previous log<br />
selection is then displayed on the second line of the screen.<br />
LOG MODE<br />
Logging=VOC, Time=05s,<br />
Max<br />
1=None 2=Auto<br />
3=VOC/FE 4=Custom<br />
Press 3=VOC/FE, then 2=FE to display<br />
this message.<br />
Enter Sampling:<br />
FE,Time=00s, Max<br />
Up/Dn=Next Type<br />
Enter=ACCEPT<br />
2. Press the Up/Dn arrow keys to choose the type of sampling to be used (Max, Avg,<br />
Smp).<br />
a. Max = the maximum reading obtained during the log interval.<br />
b. Avg = the average reading obtained during the log interval.<br />
c. Smp = the last reading obtained during the log interval.<br />
3. Type the appropriate time interval desired for F.E. sampling.<br />
4. Press Enter, and, after a brief message of acceptance, the F.E. method of logging is<br />
now ready to be used in the RUN mode.<br />
NOTE: The allowable Sample Time range is 2-30 seconds. If a Sample Time of less<br />
than 2 seconds or greater than 30 seconds is selected, a warning message will appear<br />
and you will be require to re-enter the Sample Time.<br />
Log Custom<br />
The <strong>TVA</strong>-<strong>1000B</strong> has several pre-designed logging methods. However, if you find that you<br />
would prefer to log data in a different fashion, you can use Custom logging to create your own<br />
log type. The <strong>TVA</strong>-<strong>1000B</strong> will prompt you through a series of questions which will assist you<br />
in setting the various logging options. The end result is a logging method which collects data<br />
the way you want it to.<br />