MI 611-185 – June 1996<br />

Display Menus<br />

In the LOG VOC mode, data is logged in the format described below:<br />

1. Every time RUN mode is entered a new header is created.<br />

2. Data is logged in the following format whenever you press the LOG button:<br />

Date Time PID (PPM) Alarms FID (PPM) Alarms Tag<br />

7/28/93 11:20:57:00 187.1 L 157.1 L TAG1<br />

7/28/93 11:21:29:00 436.6 H 519.6 L TAG2<br />

7/28/93 11:21:54:00 708.4 LH 847.0 LH TAG3<br />

7/28/93 11:22:19:00 300.7 L 381.9 L TAG4<br />

7/28/93 11:22:36:00 223.8 265.1 TAG5<br />

Approximately 4500 data samples may be taken.<br />

Log F.E.<br />

1. From the LOG MODE menu, press 3=VOC/FE, then 2=FE The previous log<br />

selection is then displayed on the second line of the screen.<br />

LOG MODE<br />

Logging=VOC, Time=05s,<br />

Max<br />

1=None 2=Auto<br />

3=VOC/FE 4=Custom<br />

Press 3=VOC/FE, then 2=FE to display<br />

this message.<br />

Enter Sampling:<br />

FE,Time=00s, Max<br />

Up/Dn=Next Type<br />

Enter=ACCEPT<br />

2. Press the Up/Dn arrow keys to choose the type of sampling to be used (Max, Avg,<br />

Smp).<br />

a. Max = the maximum reading obtained during the log interval.<br />

b. Avg = the average reading obtained during the log interval.<br />

c. Smp = the last reading obtained during the log interval.<br />

3. Type the appropriate time interval desired for F.E. sampling.<br />

4. Press Enter, and, after a brief message of acceptance, the F.E. method of logging is<br />

now ready to be used in the RUN mode.<br />

NOTE: The allowable Sample Time range is 2-30 seconds. If a Sample Time of less<br />

than 2 seconds or greater than 30 seconds is selected, a warning message will appear<br />

and you will be require to re-enter the Sample Time.<br />

Log Custom<br />

The <strong>TVA</strong>-<strong>1000B</strong> has several pre-designed logging methods. However, if you find that you<br />

would prefer to log data in a different fashion, you can use Custom logging to create your own<br />

log type. The <strong>TVA</strong>-<strong>1000B</strong> will prompt you through a series of questions which will assist you<br />

in setting the various logging options. The end result is a logging method which collects data<br />

the way you want it to.<br />