MI 611-185 – June 1996<br />

Theory of Operation<br />

AMPLIFIER<br />

PID 00000 PPM<br />

FID 00000 PPM<br />

%<br />

METER<br />

<strong>TVA</strong>-<strong>1000B</strong><br />

PID<br />

FID<br />

EXHAUST<br />

COLLECTOR<br />

ELECTRODE<br />

FLAME ARRESTOR<br />

LAMP<br />

JET<br />

IGNITION<br />

FILAMENT<br />

VENT<br />

AIR SAMPLE OUT<br />

COLLECTOR<br />

ELECTRODE<br />

AIR SAMPLE IN<br />

AIR SAMPLE IN<br />

HYDROGEN SUPPLY<br />

Figure 4. <strong>TVA</strong>-<strong>1000B</strong> Dual Detector Configuration<br />

Concentration Calculation and Calibration<br />

The concentration calculation is defined as the process that transforms a detector count into a<br />

displayed final concentration. The process is as follows:<br />

DETECTOR COUNT TRANSLATED INTO A STANDARD DETECTOR CONCENTRATION<br />

USING A DETECTOR COUNT VS. DETECTOR CONCENTRATION CALIBRATION CURVE.<br />

STANDARD DETECTOR CONCENTRATION IS CORRECTED TO A SPECIFIC GAS USING<br />

THE INSTRUMENT RESPONSE FACTOR.<br />

SPECIFIC GAS CONCENTRATION IS BACKGROUND CORRECTED (IF ENABLED)<br />

INSTRUMENT RESPONSE FACTOR AND BACKGROUND CORRECTED CONCENTRA-<br />

TION IS RF CORRECTED USING THE CURRENT ROUTE ENTRY RF (IF ANY)<br />

FINAL CONCENTRATION READING<br />

10