MI 611-185 – June 1996<br />

Display Menus<br />

changed. You can scroll through the other response factors by using the Up and Down arrow<br />

keys. The other response factor names will show as you scroll through them:<br />

RF1: Hexane<br />

Up/Dn=Next RF<br />

1=Change<br />

Enter=Accept<br />

If you would like to activate a different response factor, press ENTER=Accept. The<br />

<strong>TVA</strong>-1000 will show an ACCEPTED message. If you would like to modify the response factor<br />

name or numbers, press 1=Change:<br />

FID only PID only FID/PID<br />

RF1: Hexane<br />

RF1: Hexane<br />

RF1: Hexane<br />

PID: 1.00<br />

FID: 1.00<br />

PID: 1.00<br />

FID: 1.00<br />

1=Name 2=Numbers<br />

1=Name 2=Numbers<br />

1=Name 2=Numbers<br />

If you would like to change the name assigned to the response factor, press 1=Name. If you<br />

would like to change the response factor value, press 2=Numbers. If you choose 1=Name, the<br />

following display will appear:<br />

Enter name:<br />

RF1: -------------<br />

Press char key<br />

Use the alphanumeric keypad to enter a compound name for the response factor. Press enter<br />

when you are finished.<br />

If you choose 2=Numbers, the following display will appear:<br />

FID only PID only FID/PID<br />

RF1: Hexane<br />

RF1: Hexane<br />

RF1: Hexane<br />

PID: 00.00<br />

FID: 00.00<br />

P&F: 00.00<br />

Enter=Accept<br />

Enter=Accept<br />

Enter=Accept<br />

Use the keypad to enter a new response factor and press ENTER to store it in instrument memory.<br />

NOTES:<br />

1. The <strong>TVA</strong>-<strong>1000B</strong> will not accept a response factor of 00.00.<br />

2. If a response curve is used, the <strong>TVA</strong>-<strong>1000B</strong> will prompt you to enter two numbers.<br />