HIGH f<br />

When the Total Breath Rate (f) exceeds the high breath rate and time period alarm values, the HIGH<br />

f alarm is generated.<br />

• To prevent nuisance alarms, the HIGH f alarm is suspended for the first 60 seconds of ventilator<br />

operation after power up and passing the Power On Self Tests.<br />

When a HIGH f alarm occurs:<br />

• The HIGH f message is flashed in the display window.<br />

• The audible alarm is sounded.<br />

To reset the HIGH f alarm:<br />

1) Push the Silence Reset button once to silence the audible alarm.<br />

2) Push the Silence Reset button twice to reset the alarm (silences the audible alarm and clears the<br />

flashing display).<br />

• When the HIGH f alarm is reset (Silence Reset button pushed twice), the alarm is<br />

suspended for the next 60 seconds.<br />

• The 60-second suspension of the HIGH f alarm is only enabled when the alarm is manually<br />

silenced/reset by pushing the Silence Reset button twice. It is not enabled when the HIGH f<br />

alarm is automatically silenced/reset because the patient’s breath rate no longer exceeds the<br />

set HIGH f alarm value.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 9-11<br />

p/n 18247-001, Rev. G