When the Spontaneous Breathing Trial (SBT) mode of ventilation is on and the Total Breath Rate ( f )<br />

is greater than the SBT HIGH f setting (high limit of breath rate range) for 30 seconds, the SBT > f<br />

alarm is generated.<br />

For additional information concerning the Spontaneous Breathing Trial (SBT) mode of ventilation, see<br />

SBT (Spontaneous Breathing Trial) Operations in Chapter 10 - Extended Features.<br />

When a SBT > f alarm is generated:<br />

• The SBT > f message is flashed in the display window.<br />

• The audible alarm is sounded.<br />

To reset the SBT > f alarm:<br />

1) Push the Silence Reset button once to silence the audible alarm.<br />

2) Push the Silence Reset button twice to reset the alarm (silences the audible alarm and clears<br />

the flashing display).<br />

• When the SBT > f alarm is reset (Silence Reset button pushed twice), the alarm is<br />

suspended for the next 60 seconds.<br />

• The 60-second suspension of the SBT > f alarm is only enabled when the alarm is<br />

manually silenced/reset by pushing the Silence Reset button twice. It is not enabled<br />

when the SBT > f alarm is automatically silenced/reset because the patient’s breath<br />

rate is no longer greater than the set SBT HIGH f alarm value.<br />

• When an SBT alarm has been active in excess of 5 minutes, the SBT mode of<br />

ventilation is terminated, the ventilator clears the alarm status, silences the audible<br />

alarm and returns to its previous mode of ventilation / settings. To remove the flashing<br />

SBT alarm message from the display window, push the Silence Reset button.<br />

Page 9-28 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G