Event Codes<br />

This section includes a list of the event codes that can be recorded in the Event Trace.<br />

Event Codes by Code #<br />

Code Event Name Event Associated Alarm<br />

01 VENT 1 Power on None<br />

02 VENT 0 Power off None<br />

03 HOUR MTR Set hour meter None<br />

04 VENT CHK Set vent check Entered VENT CHECK mode<br />

05 APNEA 1 Apnea mode entered APNEA<br />

06 APNEA 0 Apnea mode exited APNEA<br />

07 N/A<br />

08 HIGH DIS High side disconnect DISC/SENSE<br />

09 LOW DIS Low side disconnect DISC/SENSE<br />

10 DISC 0 Circuit disconnect exited DISC/SENSE<br />

11 BATMPT1 Internal battery empty occurred BAT EMPTY<br />

12 BATMPT0 Internal battery empty exited BAT EMPTY<br />

13 BATLOW1 Internal battery low occurred BAT LOW<br />

14 BATLOW0 Internal battery low exited BAT LOW<br />

15 EXT LST1 External power lost occurred POWER LOST<br />

16 EXT LST0 External power lost exited POWER LOST<br />

17 EXT LOW1 External power low occurred POWER LOW<br />

18 EXT LOW0 External power low exited POWER LOW<br />

19 XDC FLT1 XDCR fault occurred XDCR FAULT<br />

20 XDC FLT0 XDCR fault exited XDCR FAULT<br />

21 O2 LOW 1 O2 pressure low occurred LOW O2 PRES<br />

22 O2 LOW 0 O2 pressure low exited LOW O2 PRES<br />

23 O2 HI 1 O2 pressure high occurred HIGH O2 PRES<br />

24 O2 HI 0 O2 pressure high exited HIGH O2 PRES<br />

25 DEFAULTS Defaults, or Set Defaults occurred DEFAULTS / DEFAULTS SET<br />

26 NO CAL No calibration data found NO CAL DATA<br />

27 FAN FLT1 Fan fault occurred HW FAULT<br />

28 FAN FLT0 Fan fault exited HW FAULT<br />

29 N/A<br />

30 N/A<br />

31 INTRRPT1 Spurious interrupt occurred ms RESET 1<br />

32 INTRRPT2 Spurious interrupt occurred ls RESET 1<br />

33 AD MMTCH ADC mismatch HW FAULT<br />

34 AD MTCH1 ADC mismatch occurred HW FAULT<br />

35 AD MTCH0 ADC mismatch cleared HW FAULT<br />

36 SYNCER1 Stepper motor lost sync occurred HW FAULT<br />