Alarms<br />

APNEA, APNEA xx bpm<br />

When the time since the start of the last breath is longer than the set Apnea Interval, the APNEA<br />

alarm is generated. When an Apnea alarm occurs, the ventilator will enter Apnea Backup ventilation<br />

mode. For more information on Apnea Backup mode, see Chapter 4 - Ventilation Modes, Apnea<br />

Backup. For more information on the variable Apnea Interval, see Chapter 10 - Extended Features.<br />

When an APNEA alarm occurs:<br />

• Any inspiration in progress is terminated.<br />

• The ventilator changes to Apnea Backup ventilation.<br />

• The APNEA xx bpm backup ventilation breath rate is displayed.<br />

• Control displays used while in Apnea Backup mode are illuminated and all other control displays<br />

are dimmed.<br />

• The audible alarm is sounded.<br />

While in Apnea Backup mode, the alarm will continue to sound and the alarm message and breath<br />

rate will be flashed in the display window. Apnea backup mode will continue until the operator resets<br />

the alarm or the patient triggers 2 consecutive breaths.<br />

When the APNEA alarm is reset by 2 consecutive triggered breaths:<br />

• Apnea Backup Ventilation terminates and the ventilator returns to the previous mode.<br />

• The APNEA alarm message remains flashing in the window but the breath rate is no longer<br />

displayed.<br />

• Control displays used in the selected ventilation mode are illuminated and all other control<br />

displays are dimmed.<br />

• The audible alarm is silenced.<br />

To reset the APNEA alarm and exit Apnea Backup ventilation:<br />

1) Push the Silence Reset button twice.<br />

Page 9-2 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G