WARNING<br />

Ventilation Variables and O2 Consumption - Variations in the patient’s minute ventilation, I:E ratio<br />

and/or ventilator setting changes or equipment status (i.e. circuit leaks) affect the consumption rate of<br />

oxygen. When warranted by a patients condition, it is recommended that a back-up cylinder or<br />

alternative source of oxygen be available at all times.<br />

Before Using Automobile Cigarette Lighter or Power Outlets - Before using Automobile Cigarette<br />

Lighter or Power Outlets as a power source for the <strong>LTV</strong> ® <strong>1200</strong>, assure that the ventilator’s internal<br />

battery is in good condition and fully charged. Poor cigarette lighter or power outlet connections,<br />

electrical system defects (battery, charging system, etc.), or use of vehicle accessories (air<br />

conditioner, high current lights, high power audio equipment, etc.) could result in less than the<br />

required voltage being delivered to the ventilator, generate a POWER LOST alarm and switch the<br />

ventilator’s power source to the internal battery.<br />

Unauthorized Parts or Accessories – Serious harm to the patient may result from the use of<br />

unauthorized parts or accessories. Only items expressly approved by Pulmonetic Systems may be<br />

used in conjunction with the <strong>LTV</strong> ® <strong>1200</strong> ventilators.<br />

Unapproved Adapters – Only Pulmonetic Systems Accessories should be used to connect the<br />

ventilator to Patient Assist Call Systems. These accessories incorporate safety features to reduce the<br />

risk of shock. Do not attempt to modify these accessories in any way.<br />

Patient Assist Call Connector – Do not apply more than 25V rms or 32VDC to the Patient Assist<br />

Call connector.<br />

Ventilator Service and Repair - All servicing or repair of the <strong>LTV</strong> ® <strong>1200</strong> ventilator must be performed<br />

only by a service technician certified by Pulmonetic Systems.<br />

Patient Circuits – Pulmonetic Systems Patient Circuits, Exhalation Valve Assemblies and Water<br />

Traps are shipped clean, not sterile.<br />

Ultra Violet Light Sensitivity – The material used in the tubing of the Reusable Patient Circuits is<br />

not UV stable. Avoid exposure of the tubing to UV light.<br />

Mounting Screws - Refer to the information contained in Pulmonetic Systems Replacement Screws<br />

Kit, P/N 11149, to determine the appropriate accessories mounting screws or accessories<br />

replacement screws location, type and length to use when removing or exchanging external<br />

accessories on an <strong>LTV</strong> ® ventilator.<br />

Mounting Screw Use – Internal damage to the ventilator may result if the wrong length mounting<br />

screws are used when installing or removing external accessories.<br />

Patient Circuit Accessories - The use of accessories such as Speaking Valves, Heat-Moisture<br />

Exchangers and Filters create additional patient circuit resistance and in the event of a disconnection,<br />

may impede the generation of a Low Pressure alarm. Ensure that the Low Pressure alarm settings<br />

accommodate these types of accessories when used in combination with patient circuits.<br />

Low Minute Volume Control Settings - The Low Min. Vol. control should be set to its highest<br />

clinically appropriate value. If there is a clinical need to set the Low Minute Volume alarm to lower<br />

values or off (“- - -“), perform a clinical assessment to determine if an alternative monitor (i.e. a Pulse<br />

Oxymeter with an audible alarm, or a Cardio Respiratory Monitor) should be used.<br />

The MR Conditional <strong>LTV</strong> ® <strong>1200</strong> System is specified as - An <strong>LTV</strong> ® <strong>1200</strong> Ventilator (P/N 18888-<br />

2XX), <strong>LTV</strong> ® <strong>1200</strong> MR Safe 15ft Patient Circuit (P/N 19189-001), MR Conditional Floor Stand (P/N<br />

14982-001), and a <strong>LTV</strong> ® AC adapter (P/N 18053-001). Adding not approved parts or accessories may<br />

cause patient and/or operator harm.<br />