• Item (5), #4-40 X 1/4” Flat-head mounting screw (1) P/N 10430 76<br />

• Item (6), Finish Washers (8) P/N 10191, 19119-001, or 19119-002 76<br />

• Item (7), #4-40 X 1/2” Flat-head mounting screws (6) P/N 10338 76<br />

• Item (8), #4-40 X 3/8” Flat-head mounting screws (2) P/N 10474 76<br />

5) Insert and thread two #4-40 X 1/2” flat-head mounting screws (7) with finish washers (6) through<br />

the screw holes in the sides of the upper boot, as indicated in the illustration.<br />

6) Torque tighten the mounting screws to these specified values (do not over tighten to avoid<br />

damage to the finish washers).<br />

• Torque tighten the screws in the legs of the boot to 60 in-oz (0.42 Nm)<br />

• Torque tighten the screws in the sides of the boot to 20 in-oz (0.14 Nm)<br />

76 Contained in Pulmonetic Systems Replacement Screws kit, P/N 11149.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page C-7<br />

p/n 18247-001, Rev. G