Variable Rise Time<br />

Use the Variable Rise Time option to select the rise time profile for Pressure Control and Pressure<br />

Support breaths. The rise time profiles are numbered 1 through 9 where 1 is the fastest rise time and<br />

9 is the slowest rise time. Starting with the fastest rise time, each time is 33% longer than the<br />

previous one.<br />

To modify the Rise Time Profile:<br />

1) Push the Select button while RISE TIME is displayed.<br />

2) PROFILE x is displayed.<br />

3) Turn the Set Value knob until the desired Rise Time Profile is displayed.<br />

4) Push the Select button.<br />

Range: 1 to 9, where 1 is the fastest and 9 is the slowest<br />

Page 10-8 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G