HIGH O 2 PRES<br />

When the average oxygen inlet pressure exceeds the acceptable limit for the type of oxygen source,<br />

the HIGH O2 PRES alarm is generated.<br />

• If Low Pressure O2 Source is selected, the inlet pressure is >10 PSIG.<br />

• If Low Pressure O2 Source is not selected and the oxygen concentration is set to greater than<br />

21%, the inlet pressure is >85 PSIG.<br />

When the Low Pressure O2 Source option is selected and a high O2 pressure source is attached to<br />

the ventilator, an Automatic High O2 Switch Over safety response generates a HIGH O2 PRES alarm,<br />

switches the ventilator to High Pressure O2 Source mode and sets the percentage of oxygen to be<br />

delivered in the gas flow to 21%.<br />

When a HIGH O2 PRES alarm occurs:<br />

• The HIGH O2 PRES message is flashed in the display window.<br />

• The O2 % (O2 Flush) control display is flashed.<br />

• The audible alarm is sounded.<br />

To reset the HIGH O2 PRES alarm:<br />

1) Push the Silence Reset button to silence the alarm.<br />

2) Adjust the oxygen inlet pressure.<br />

3) Push the Silence Reset button to reset the alarm.<br />

Page 9-12 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />