Bright, Dim and Blank Control Displays<br />

Variable controls will be displayed at normal or dimmed intensity, or may be blanked. A display will be<br />

displayed at normal intensity:<br />

• When it is selected for change. All other displays will be dimmed.<br />

• When it is active in the current ventilation mode. Dimmed displays are not active in the current<br />

mode.<br />

NOTE<br />

Be sure to set any controls that may be used in Apnea Backup ventilation to appropriate values. Even<br />

though these controls are dimmed, they will be used if apnea should occur.<br />

REMARQUE<br />

Assurez-vous de régler aux valeurs appropriées, tous les contrôles susceptibles d'être utilisés en<br />

mode ventilation de secours pour l'apnée. Même si ces contrôles sont en veilleuse, ils seront utilisés<br />

en cas d'apnée.<br />

A display will be displayed at dimmed intensity:<br />

• When another control is selected for change.<br />

• When it is not active in the current ventilation mode.<br />

A display will be blank:<br />

• To conserve battery power while operating from battery power:<br />

• If no button pushes or control knob activity occurs for 60 seconds, the displays are turned off.<br />

The display window, 7-segment control displays, and LEDs are turned off. Anytime an alarm<br />

occurs, or if an alarm message is already displayed, the display window will remain active.<br />

The Airway Pressure display is always active.<br />

• To turn the displays back on, push any button or turn the control knob.<br />

• When an option, such as oxygen blending, is not installed.<br />

• When a control feature is not available, such as during Ventilator Checkout tests.<br />

Page 5-4 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />