DEFAULTS (cont.)<br />

The factory-set default Control settings are:<br />

• See Chapter 6 - Controls, for detailed information concerning specific Controls.<br />

Control - Default Control - Default<br />

Breath Rate - 12 bpm O2 % (O2 Flush) - 21%<br />

Control Lock - On PEEP - 0 cmH2O<br />

Data Display Scrolling - Auto-On Pressure Control - 1 cmH2O<br />

High Pressure Limit - 20 cmH2O Pressure Support - 1 cmH2O<br />

Inspiratory/Expiratory Hold - Off Sensitivity - 2 Lpm<br />

Inspiratory Time - 1.5 sec Tidal Volume - 500 ml<br />

Low Minute Volume - 2.5 Lpm Ventilation Mode - Assist/Control<br />

Low Pressure - 5 cmH2O<br />

Low Pressure O2 Source - Off<br />

Volume Pressure Mode - Volume<br />

The factory-set default Extended Features settings are:<br />

• See Chapter 10 - Extended Features, for detailed information concerning specific Features.<br />

Feature - Default Feature - Default<br />

Alarm Volume - 85 dBA Patient Assist - Normal<br />

Apnea Interval - 20 sec Patient Query - On<br />

Com Setting 38 - Data Patient Size - INFANT<br />

Control Unlock - Easy PC Flow Termination - Off<br />

Date Format - MM/DD/YYYY PIP LED - On<br />

High f Alarm - High f Off Rise Time Profile - 4<br />

High f Alarm Delay - 30 sec SBT Display f/Vt Alarm - Off<br />

High PEEP Alarm - PEEP +5 cmH2O SBT FIO2 - 21%<br />

HP Alarm Delay - No Delay SBT High f Alarm - 35 bpm<br />

Language - English SBT High f/Vt Alarm - 105 f/Vt<br />

Leak Compensation - On SBT Low f Alarm - 10 bpm<br />

Low PEEP Alarm - PEEP -5 cmH2O SBT Low f/Vt Alarm - 70 f/Vt<br />

LPP Alarm - All Breaths SBT Mode - Off<br />

NPPV Mode - Off SBT Mode Run Time - 20 min<br />

O2 Conserve - Off SBT PEEP - 0 cmH2O<br />

O2 Cylinder Duration - 2000 psi or 138 bar SBT Pressure Support - 10 cmH2O<br />

O2 Cylinder Duration - 622 liters Variable Flow Termination - 25%<br />

O2 Flush Period - 3 min Variable Time Termination - 1.5 sec<br />

38 This feature is not reset to default values when the SET DEFAULTS option is used in Extended Features.<br />

Page 9-8 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />