28.02.2013 Views

LTV 1200 Operator's Manual - CareFusion Emergency Preparedness

LTV 1200 Operator's Manual - CareFusion Emergency Preparedness

LTV 1200 Operator's Manual - CareFusion Emergency Preparedness

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Fixed Alarms<br />

Control Range Tolerance / Indicators<br />

Default Settings EEPROM problem detected n/a<br />

DISC/SENSE<br />

Positive (exhaled) airway flow n/a<br />

(Low Pressure Sense<br />

Line Disconnect)<br />

during first 200 ms of<br />

inspiration and exhaled tidal<br />

volume (Vte) of previous<br />

breath is more than 4000 ml<br />

DISC/SENSE<br />

(High Pressure Sense<br />

Line Disconnect)<br />

Airway pressure changes by <<br />

1 cmH2O during 200 ms after<br />

inspiratory start<br />

OR<br />

After initial 200 ms of<br />

inspiration airway pressure<br />

drops below 0.125 cmH2O and<br />

can’t be raised more than 0.5<br />

cmH2O in next 500 ms<br />

External Power Lost

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!