LMV OFF<br />

The LMV OFF message is displayed when;<br />

• The Low Minute Volume alarm is turned off by being set to dashes and,<br />

• The LPP alarm is set to ALL BREATHS.<br />

This is an informational message only. The message is displayed at power up, when monitored data<br />

is being scrolled automatically, and when no front panel activity is detected for 60 seconds.<br />

To clear the LMV OFF message:<br />

1) Push any front panel button or turn the Set Value knob.<br />

LO PEEP OFF<br />

The LO PEEP OFF message is displayed when;<br />

• The Low PEEP alarm is turned off by being set to LO PEEP OFF and,<br />

• The High Breath Rate alarm (HIGH f) is turned on/set to a value.<br />

This is an informational message only. The message is displayed at power up, when monitored data<br />

is being scrolled automatically, and when no front panel activity is detected for 60 seconds.<br />

To clear the LO PEEP OFF message:<br />

1) Push any front panel button or turn the Set Value knob.<br />

LOCKED<br />

The LOCKED message is displayed when a button is pushed while the controls are locked. No<br />

audible alarm is given.<br />

When a LOCKED message is displayed:<br />

• The LOCKED message is flashed in the display window for 5 seconds or until the controls<br />

are unlocked.<br />

• Control settings may not be changed.<br />

There are two methods for unlocking the controls: EASY and HARD. The unlock method is selected<br />

under the Extended Features menus 47 .<br />

To unlock the controls with EASY unlocking:<br />

1) Push the Control Lock button.<br />

To unlock the controls with HARD unlocking:<br />

1) Push and hold the Control Lock button for 3 seconds.<br />

47 See Chapter 10 - Extended Features, Control Unlock for more information.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 9-35<br />

p/n 18247-001, Rev. G