Problem Possible Causes What To Do<br />

Battery doesn’t reach<br />

full charge.<br />

Battery depletes too<br />

quickly.<br />

Battery Charge<br />

Status LED is<br />

flashing amber.<br />

Battery Charge<br />

Status LED is<br />

flashing red.<br />

Battery Charge<br />

Status LED is solid<br />

red.<br />

Internal battery deeply<br />

discharged.<br />

Defective internal<br />

battery or internal<br />

problem with the<br />

ventilator.<br />

Internal battery<br />

charging.<br />

Defective internal<br />

battery or internal<br />

problem with the<br />

ventilator.<br />

Defective internal<br />

battery or internal<br />

problem with the<br />

ventilator.<br />

Defective internal<br />

battery or internal<br />

problem with the<br />

ventilator.<br />

Charge the internal battery for 24 hours by<br />

connecting the external AC adapter and plugging<br />

it into a good AC source. If the battery is deeply<br />

discharged, it may take several cycles of<br />

charging and discharging for the battery to reach<br />

a maximum charge.<br />

Immediately contact a certified Pulmonetic<br />

Systems’ service technician.<br />

The Charge Status LED flashes amber while the<br />

battery charging circuit evaluates the battery as a<br />

part of the charge cycle. If the battery is found to<br />

be OK, the Charge Status LED will change to<br />

solid amber while the battery is charging. The<br />

internal battery charges any time the ventilator is<br />

connected to an external power source. If the<br />

battery is deeply discharged, the Charge Status<br />

LED may flash amber for up to an hour.<br />

Immediately contact a certified Pulmonetic<br />

Systems’ service technician.<br />

Immediately contact a certified Pulmonetic<br />

Systems’ service technician.<br />

Immediately contact a certified Pulmonetic<br />

Systems’ service technician.<br />